期刊论文详细信息
Journal of Cardiothoracic Surgery | |
Pacemaker lead malposition in the left atrial roof is masked by normal pacing thresholds | |
Alfred Kocher1  Andreas Habertheuer1  Dominik Wiedemann1  Caesar Khazen1  Martin Andreas1  Claus Rath1  | |
[1] Division of Cardiac Surgery, Vienna General Hospital, Medical University of Vienna, Waehringer Guertel 18-20, Vienna A-1090, Austria | |
关键词: Chest X-ray; Cardiac tamponade; Lead perforation; Lead malpositioning; Pacemaker; | |
Others : 811202 DOI : 10.1186/1749-8090-9-53 |
|
received in 2013-05-17, accepted in 2014-03-11, 发布年份 2014 | |
【 摘 要 】
Pacemaker lead malpositioning with subsequent cardiac tamponade is a rare, but serious adverse event. We herein report a case of pacemaker lead malpositioning in a 76-year old female caucasian patient. The lead was malpositioned into the roof of the left atrium after perforation of the superior vena cava, resulting in cardiac tamponade. After fast surgical revision and an uneventful post-operative period, the patient was discharged in excellent condition.
【 授权许可】
2014 Rath et al.; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20140709061300746.pdf | 1643KB | download | |
Figure 3. | 30KB | Image | download |
Figure 2. | 103KB | Image | download |
Figure 1. | 50KB | Image | download |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
【 参考文献 】
- [1]Palmisano P, Accogli M, Zaccaria M, Luzzi G, Nacci F, Anaclerio M, Favale S: Rate, causes, and impact on patient outcome of implantable device complications requiring surgical revision: large population survey from two centres in Italy. Europace 2013, 15:531-540.
- [2]Udo EO, Zuithoff NP, van Hemel NM, de Cock CC, Hendriks T, Doevendans PA, Moons KG: Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm 2012, 9:728-735.
- [3]Rajappan K: Permanent pacemaker implantation technique: part II. Heart 2009, 95:334-342.
- [4]Lelorier P: Accidents will happen (so be prepared). Heart Rhythm 2005, 2:912-913.
- [5]Mahapatra S, Bybee KA, Bunch TJ, Espinosa RE, Sinak LJ, McGoon MD, Hayes DL: Incidence and predictors of cardiac perforation after permanent pacemaker placement. Heart Rhythm 2005, 2:907-911.
- [6]Armaganijan LV, Toff WD, Nielsen JC, Andersen HR, Connolly SJ, Ellenbogen KA, Healey JS: Are elderly patients at increased risk of complications following pacemaker implantation? A meta-analysis of randomized trials. Pacing Clin Electrophysiol 2012, 35:131-134.
- [7]Robinson JF, Robinson WA, Cohn A, Garg K, Armstrong JD 2nd: Perforation of the great vessels during central venous line placement. Arch Intern Med 1995, 155:1225-1228.
- [8]Fann JI, Atwood JE, Burdon TA: Repair of superior vena caval perforation during pacemaker placement with video-assisted limited thoracotomy. Heart Surg Forum 2004, 7:1-3.
- [9]Costelloe CM, Murphy WA Jr, Gladish GW, Rozner MA: Radiography of pacemakers and implantable cardioverter defibrillators. AJR Am J Roentgenol 2012, 199:1252-1258.
- [10]Edwards NC, Varma M, Pitcher DW: Routine chest radiography after permanent pacemaker implantation: is it necessary? J Postgrad Med 2005, 51:92-96. discussion 96–97