期刊论文详细信息
Journal of Orthopaedic Surgery and Research
An intraoperative device to restore femoral offset in total hip arthroplasty
Yu Zhang2  Hong Wen2  Paul Kim Chiu Wong3  Chengwang Chen2  Zhen Su1  Enxing Xue2 
[1] Department of Nephrology, The First Affiliated Hospital of Wenzhou Medical University, Wenzhou 325000, Zhejiang, China;Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children’s Hospital of Wenzhou Medical University, 109 Xueyuan West Road, Wenzhou 325000, Zhejiang, China;Department of Surgery, University of Toronto, 650 Sammon Ave. Suite 301, Toronto M4C 5M5, ON, Canada
关键词: Device;    Femoral offset;    Total hip arthroplasty;    Leg length discrepancy;   
Others  :  1152349
DOI  :  10.1186/s13018-014-0058-7
 received in 2013-11-25, accepted in 2014-06-27,  发布年份 2014
PDF
【 摘 要 】

Background

Leg length discrepancy (LLD) after total hip arthroplasty (THA) can lead to unsatisfactory outcome. Our objective was to design and evaluate a simple and reliable intraoperative device (Length-offset Lever) to minimize leg length discrepancy.

Methods

This device was used in 51 patients undergoing primary total hip replacements. The leg length discrepancy was measured pre- and postoperatively based on plain radiographs.

Results

Preoperative radiographic leg length discrepancy averaged 13.5 ± 6.2 mm. Leg length discrepancy showed significant improvement, with a postoperative average of 4.1 ± 2.3 mm (p < 0.0001). There were no complications associated with this device.

Conclusions

The ‘Length-offset Lever’ is a useful intraoperative tool to restore anatomic femoral offset and height of femoral head.

【 授权许可】

   
2014 Xue et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150406162607397.pdf 3051KB PDF download
Figure 4. 27KB Image download
Figure 3. 29KB Image download
Figure 2. 38KB Image download
Figure 1. 46KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

【 参考文献 】
  • [1]Charnley J: Low Friction Arthroplasty of the Hip: Theory and Practice. Springer-Verlag, New York; 1979.
  • [2]Konyves A, Bannister GC: The importance of leg length discrepancy after total hip arthroplasty. J Bone Joint Surg (Br) 2005, 87:155.
  • [3]Yamaguchi T, Naito M, Aasayama I, Ishiko T: Total hip arthroplasty: the relationship between posterolateral reconstruction, abductor muscle strength, and femoral offset. J Orthop Surg (Hong Kong) 2004, 12:164.
  • [4]McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME: Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg (Br) 1995, 77-B:865.
  • [5]Asayama I, Naito M, Fujisawa M, Kambe T: Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty 2002, 17:747.
  • [6]Asayama I, Chammongkich S, Simpson KJ, Kinsey TL, Mahoney OM: Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 2005, 20:414.
  • [7]Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH: Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop 2001, 388:125.
  • [8]Naito M, Ogata K, Asayama I: Intraoperative limb length measurement in total hip arthroplasty. Int Orthop 1999, 23:31.
  • [9]Ranawat CS, Rajesh R, Rodriquez JA, Bhende HS: Correction of limb length inequality during total hip arthroplasty. J Arthroplasty 2001, 16:715.
  • [10]Shiramizu K, Naito M, Shitama T, Nakamura Y, Shitama H: L-shaped caliper for limb length measurement during total hip arthroplasty. J Bone Joint Surg (Br) 2004, 86:966.
  • [11]Takigami I, Itokazu M, Itoh Y, Matsumoto K, Yamamoto T, Shimizu K: Limb-length measurement in total hip arthroplasty using a calipers dual pin retractor. Bull NYU Hosp Jt Dis 2008, 66(2):107.
  • [12]Harris VM: Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969, 51:737.
  • [13]Kelley SS: High hip center in revision arthroplasty. J Arthroplasty 1994, 9:503.
  • [14]Charles MN, Bourne RB, Davey JR, Greenwald AS, Morrey BF, Rorabeck CH: Soft-tissue balancing of the hip: the role of femoral offset restoration. Instr Course Lect 2005, 54:131.
  • [15]Bourne RB, Rorabeck CH, Patterson JJ, Guerin J: Tapered titanium cementless total hip replacements: a 10- to 13-year follow-up study. Clin Orthop 2001, 393:112.
  • [16]Konyves A, Bannister GC: The importance of leg length discrepancy after total hiparthroplasty. J Bone Joint Surg (Br) 2005, 87-B:155.
  • [17]Steinberg B, Harris WH: The offset problem in total hip arthroplasty. Contemp Orthop 1992, 24:556.
  • [18]Plaass C, Clauss M, Ochsner PE, Ilchmann T: Influence of leg length discrepancy on clinical results after total hip arthroplasty: a prospective clinical trial. Hip Int 2011, 21:441.
  • [19]Tannast M, Zheng G, Anderegg C, Burckhardt K, Langlotz F, Ganz R, Siebenrock KA: Tilt and rotation correction of acetabular version on pelvic radiographs. Clin Orthop Relat Res 2005, 438:182.
  • [20]McGee HMJ, Scott JHS: A simple method of obtaining equal leg length in total hip arthroplasty. Clin Orthop Relat Res 1985, 194:207.
  • [21]Itokazu M, Masuda K, Ohno T, Itoh Y, Takatsu T, Wenyi Y: A simple method of intraoperative limb length measurement in total hip arthroplasty. Bull Hosp Jt Dis 1997, 56:204.
  • [22]Sathappan SS, Ginat D, Patel V, Walsh M, Jaffe WL, Di Cesare PE: Effect of anesthesia type on limb length discrepancy after total hip arthroplasty. J Arthroplasty 2008, 23(2):203-209.
  • [23]Ng VY, Kean JR, Glassman AH: Limb-length discrepancy after hip arthroplasty. J Bone Joint Surg Am 2013, 95(15):1426-1436.
  文献评价指标  
  下载次数:83次 浏览次数:31次