[1] Department of Nephrology, The First Affiliated Hospital of Wenzhou Medical University, Wenzhou 325000, Zhejiang, China;Department of Orthopaedic Surgery, The Second Affiliated Hospital and Yuying Children’s Hospital of Wenzhou Medical University, 109 Xueyuan West Road, Wenzhou 325000, Zhejiang, China;Department of Surgery, University of Toronto, 650 Sammon Ave. Suite 301, Toronto M4C 5M5, ON, Canada | |
关键词: Device; Femoral offset; Total hip arthroplasty; Leg length discrepancy; | |
Others : 1152349 DOI : 10.1186/s13018-014-0058-7 |
|
received in 2013-11-25, accepted in 2014-06-27, 发布年份 2014 | |
![]() |
【 摘 要 】
Background
Leg length discrepancy (LLD) after total hip arthroplasty (THA) can lead to unsatisfactory outcome. Our objective was to design and evaluate a simple and reliable intraoperative device (Length-offset Lever) to minimize leg length discrepancy.
Methods
This device was used in 51 patients undergoing primary total hip replacements. The leg length discrepancy was measured pre- and postoperatively based on plain radiographs.
Results
Preoperative radiographic leg length discrepancy averaged 13.5 ± 6.2 mm. Leg length discrepancy showed significant improvement, with a postoperative average of 4.1 ± 2.3 mm (p < 0.0001). There were no complications associated with this device.
Conclusions
The ‘Length-offset Lever’ is a useful intraoperative tool to restore anatomic femoral offset and height of femoral head.
【 授权许可】
2014 Xue et al.; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20150406162607397.pdf | 3051KB | ![]() |
|
Figure 4. | 27KB | Image | ![]() |
Figure 3. | 29KB | Image | ![]() |
Figure 2. | 38KB | Image | ![]() |
Figure 1. | 46KB | Image | ![]() |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
【 参考文献 】
- [1]Charnley J: Low Friction Arthroplasty of the Hip: Theory and Practice. Springer-Verlag, New York; 1979.
- [2]Konyves A, Bannister GC: The importance of leg length discrepancy after total hip arthroplasty. J Bone Joint Surg (Br) 2005, 87:155.
- [3]Yamaguchi T, Naito M, Aasayama I, Ishiko T: Total hip arthroplasty: the relationship between posterolateral reconstruction, abductor muscle strength, and femoral offset. J Orthop Surg (Hong Kong) 2004, 12:164.
- [4]McGrory BJ, Morrey BF, Cahalan TD, An KN, Cabanela ME: Effect of femoral offset on range of motion and abductor muscle strength after total hip arthroplasty. J Bone Joint Surg (Br) 1995, 77-B:865.
- [5]Asayama I, Naito M, Fujisawa M, Kambe T: Relationship between radiographic measurements of reconstructed hip joint position and the Trendelenburg sign. J Arthroplasty 2002, 17:747.
- [6]Asayama I, Chammongkich S, Simpson KJ, Kinsey TL, Mahoney OM: Reconstructed hip joint position and abductor muscle strength after total hip arthroplasty. J Arthroplasty 2005, 20:414.
- [7]Sakalkale DP, Sharkey PF, Eng K, Hozack WJ, Rothman RH: Effect of femoral component offset on polyethylene wear in total hip arthroplasty. Clin Orthop 2001, 388:125.
- [8]Naito M, Ogata K, Asayama I: Intraoperative limb length measurement in total hip arthroplasty. Int Orthop 1999, 23:31.
- [9]Ranawat CS, Rajesh R, Rodriquez JA, Bhende HS: Correction of limb length inequality during total hip arthroplasty. J Arthroplasty 2001, 16:715.
- [10]Shiramizu K, Naito M, Shitama T, Nakamura Y, Shitama H: L-shaped caliper for limb length measurement during total hip arthroplasty. J Bone Joint Surg (Br) 2004, 86:966.
- [11]Takigami I, Itokazu M, Itoh Y, Matsumoto K, Yamamoto T, Shimizu K: Limb-length measurement in total hip arthroplasty using a calipers dual pin retractor. Bull NYU Hosp Jt Dis 2008, 66(2):107.
- [12]Harris VM: Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J Bone Joint Surg Am 1969, 51:737.
- [13]Kelley SS: High hip center in revision arthroplasty. J Arthroplasty 1994, 9:503.
- [14]Charles MN, Bourne RB, Davey JR, Greenwald AS, Morrey BF, Rorabeck CH: Soft-tissue balancing of the hip: the role of femoral offset restoration. Instr Course Lect 2005, 54:131.
- [15]Bourne RB, Rorabeck CH, Patterson JJ, Guerin J: Tapered titanium cementless total hip replacements: a 10- to 13-year follow-up study. Clin Orthop 2001, 393:112.
- [16]Konyves A, Bannister GC: The importance of leg length discrepancy after total hiparthroplasty. J Bone Joint Surg (Br) 2005, 87-B:155.
- [17]Steinberg B, Harris WH: The offset problem in total hip arthroplasty. Contemp Orthop 1992, 24:556.
- [18]Plaass C, Clauss M, Ochsner PE, Ilchmann T: Influence of leg length discrepancy on clinical results after total hip arthroplasty: a prospective clinical trial. Hip Int 2011, 21:441.
- [19]Tannast M, Zheng G, Anderegg C, Burckhardt K, Langlotz F, Ganz R, Siebenrock KA: Tilt and rotation correction of acetabular version on pelvic radiographs. Clin Orthop Relat Res 2005, 438:182.
- [20]McGee HMJ, Scott JHS: A simple method of obtaining equal leg length in total hip arthroplasty. Clin Orthop Relat Res 1985, 194:207.
- [21]Itokazu M, Masuda K, Ohno T, Itoh Y, Takatsu T, Wenyi Y: A simple method of intraoperative limb length measurement in total hip arthroplasty. Bull Hosp Jt Dis 1997, 56:204.
- [22]Sathappan SS, Ginat D, Patel V, Walsh M, Jaffe WL, Di Cesare PE: Effect of anesthesia type on limb length discrepancy after total hip arthroplasty. J Arthroplasty 2008, 23(2):203-209.
- [23]Ng VY, Kean JR, Glassman AH: Limb-length discrepancy after hip arthroplasty. J Bone Joint Surg Am 2013, 95(15):1426-1436.