Rate of manual leukocyte differentials in dog, cat and horse blood samples using ADVIA 120 cytograms | |
Natali Bauer1  Andreas Moritz1  Martina Stirn2  | |
[1] Department of Veterinary Clinical Sciences, Clinical Pathology and Clinical Pathophysiology, Justus-Liebig University Giessen, Frankfurterstr. 126, Giessen 35392, Germany;Clinical laboratory, University of Zurich, Vetsuisse Faculty, University Zurich, Winterthurerstr. 260, CH-8057 Zurich, Switzerland | |
关键词: Horse; Cat; Dog; Reactive lymphocytes; Left shift; Laboratory productivity; Blood smear review; Cytograms; White blood cells; | |
Others : 866781 DOI : 10.1186/1746-6148-10-125 |
|
received in 2013-11-22, accepted in 2014-05-29, 发布年份 2014 | |
【 摘 要 】
Background
Modern automated haematology instruments are capable of performing leukocyte differentials faster, cheaper and with a higher precision than the traditional 100-cell manual differential count. Thus, in human laboratories, criteria are defined for performing a manual review of the blood smear resulting in a marked reduction of manual differential counts. While common in human laboratories, this approach to reducing the number of manual differentials in veterinary laboratories is still not commonly performed. Thus, our aim was to determine the rate and causes of manual leukocyte differentials in a university clinical pathology laboratory using the automated laser-based haematology analyser ADVIA 120. Overall, 14,953 complete blood cell counts from dogs, cats and horses were reviewed. Manual leukocyte differentials were requested if abnormal ADVIA peroxidase and baso cytograms were detected (i.e. suspicion of left shift or atypical lymphocytes/blasts, inappropriate separation of cell populations).
Results
In 21% of canine, 32% of feline and 20% of equine samples, a manual differential was requested. Indistinct separation of the cell population was present in 10% to 15% of the cases. Depending on the species, atypical lymphocytes were suspected in 2% to 12%, left shift in 13% to 25% and suspicion of blasts was present in less than 0.4% of the cases.
Conclusions
The obtained results are comparable to those published for human medicine and the rate of manual differentiation could be markedly reduced in veterinary laboratories if microscopic examination was used as a validation procedure rather than as a reflexive substitute for automated differentiation.
【 授权许可】
2014 Stirn et al.; licensee BioMed Central Ltd.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
20140728022212734.pdf | 1249KB | download | |
158KB | Image | download | |
168KB | Image | download | |
205KB | Image | download |
【 图 表 】
【 参考文献 】
- [1]Moritz A: Der Einsatz lasergestützter Multiparameter-Hämatologiesysteme in der Veterinärmedizin [The diagnostic use of laser based multiparameter hematology analyzers in veterinary medicine]. Wetzlar, Germany: Pandora Verlag Wetzlar; 2000.
- [2]Moritz A, Fickenscher Y, Meyer K, Failing K, Weiss DJ: Canine and feline hematology reference values for the ADVIA 120 hematology system. Vet Clin Pathol 2004, 33:32-38.
- [3]Lilliehöök I, Tvedten H: Validation of the Sysmex XT-2000iV hematology system for dogs, cats, and horses. II. Differential leukocyte counts. Vet Clin Pathol 2009, 38:175-182.
- [4]Lilliehöök I, Tvedten H: Validation of the Sysmex XT-2000iV hematology system for dogs, cats, and horses. I. Erythrocytes, platelets, and total leukocyte counts. Vet Clin Pathol 2009, 38:163-174.
- [5]Neuerer FF: Evaluation des vollautomatischen Hämatologiegerätes CELL-DYN 3500 im klinischen Einsatz bei Hund und Katze [Evaluation of the automated hematology analyzer CELL-DYN 3500 for its clinical use in cats and dogs]. Munich, Germany: Dissertation München; 2005.
- [6]Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A: Evaluation of the automated hematology analyzer Sysmex XT-2000iV compared to the ADVIA(R) 2120 for its use in dogs, cats, and horses: part I–precision, linearity, and accuracy of complete blood cell count. J Vet Diagn Invest 2011, 23:1168-1180.
- [7]Bauer N, Nakagawa J, Dunker C, Failing K, Moritz A: Evaluation of the automated hematology analyzer Sysmex XT-2000iV compared to the ADVIA(R) 2120 for its use in dogs, cats, and horses. Part II: Accuracy of leukocyte differential and reticulocyte count, impact of anticoagulant and sample aging. J Vet Diagn Invest 2012, 24:74-89.
- [8]Mathers RA, Evans GO, Bleby J, Tornow T: Evaluation of the Sysmex XT-2000iV haematology analyser for rat, dog and mouse whole blood samples. Comp Clin Pathol 2008, 17:137-144.
- [9]Goldmann F, Bauer N, Moritz A: Evaluation of the IDEXX ProCyte Dx analyzer for dogs and cats compared to the Siemens ADVIA 2120 and manual differential. Comp Clin Pathol 2012, 2303:283-296.
- [10]Goldmann F, Bauer N, Moritz A: Comparison of the ProCyte Dx analyzer with the ADVIA 2120 and the manual differential for validation of equine and bovine hemograms. Comp Clin Pathol 2012, 2205:855-868.
- [11]Novis DA, Walsh M, Wilkinson D, St Louis M, Ben Ezra J: Laboratory productivity and the rate of manual peripheral blood smear review: a College of American Pathologists Q-Probes study of 95,141 complete blood count determinations performed in 263 institutions. Arch Pathol Lab Med 2006, 130:596-601.
- [12]Rabizadeh E, Pickholtz I, Barak M, Froom P: Historical data decrease complete blood count reflex blood smear review rates without missing patients with acute leukaemia. J Clin Pathol 2013, 66:692-694.
- [13]Lantis KL, Harris RJ, Davis G, Renner N, Finn WG: Elimination of instrument-driven reflex manual differential leukocyte counts. Optimization of manual blood smear review criteria in a high-volume automated hematology laboratory. Am J Clin Pathol 2003, 119:656-662.
- [14]Tvedten HW, Haines C: Canine Automated Differential Leukocyte Count: Study Using a Hematology Analyzer System. Vet Clin Pathol 1994, 23:90-96.
- [15]Tvedten HW, Korcal D: Automated Differential Leukocyte Count in Horses, Cattle, and Cats Using the Technicon H-1E Hematology System. Vet Clin Pathol 1996, 25:14-22.
- [16]Barnes PW, McFadden SL, Machin SJ, Simson E: The international consensus group for hematology review: suggested criteria for action following automated CBC and WBC differential analysis. Lab Hematol 2005, 11:83-90.
- [17]Bain BJ: Current Concepts: Diagnosis from the Blood Smear. N Engl J Med 2005, 353:498-507.
- [18]Bayer Health Care LLC: ADVIA 120 Operator's Guide. 2004.
- [19]Harris N, Kunicka J, Kratz A: The ADVIA 2120 Hematology System: Flow Cytometry-Based Analysis of Blood and Body Fluids in the Routine Hematology Laboratory. Lab Hematol 2005, 11:47-61.
- [20]Moritz A, Hoffmann C: Platelet count in the cat. Tierarztliche Praxis 1997, 25:695-700.
- [21]Rümke CL: Variability of results in differential counts on blood smears. Triangle 1960, 44:154-158.
- [22]Clark P, Mogg TD, Tvedten HW, Korcal D: Artifactual changes in equine blood following storage, detected using the Advia 120 hematology analyzer. Vet Clin Pathol 2002, 31:90-94.
- [23]Furlanello T, Tasca S, Caldin M, Carli E, Patron C, Tranquillo M, Lubas G, Solano-Gallego L: Artifactual changes in canine blood following storage, detected using the ADVIA 120 hematology analyzer. Vet Clin Pathol 2006, 35:42-46.