International Journal of Behavioral Nutrition and Physical Activity
Mechanical and free living comparisons of four generations of the Actigraph activity monitor
Niels Christian Møller3  Lars Bo Andersen1  May Grydeland1  Bjørge Herman Hansen1  Søren Brage2  Jan Christian Brønd3  Mathias Ried-Larsen3 
[1] Department of Sports Medicine, Norwegian School of Sport Sciences, Oslo, Norway;MRC Epidemiology Unit, Institute of Metabolic Science, Cambridge, UK;University of Southern Denmark, Institute of Sports Science and Clinical Biomechanics, Centre of Research in Childhood Health, Campusvej, Odense M, 55 5230, Denmark
关键词: Free living;    Mechanical testing;    Physical activity intensity;    Physical activity assessment;    Accelerometry;   
Others  :  811788
DOI  :  10.1186/1479-5868-9-113
 received in 2012-03-21, accepted in 2012-09-04,  发布年份 2012
PDF
【 摘 要 】

Background

More studies include multiple generations of the Actigraph activity monitor. So far no studies have compared the output including the newest generation and investigated the impact on the output of the activity monitor when enabling the low frequency extension (LFE) option. The aims were to study the responses of four generations (AM7164, GT1M, GT3X and GT3X+) of the Actigraph activity monitor in a mechanical setup and a free living environment with and without enabling the LFE option.

Methods

The monitors were oscillated in a mechanical setup using two radii in the frequency range 0.25-3.0 Hz. Following the mechanical study a convenience sample (N = 20) wore three monitors (one AM7164 and two GT3X) for 24 hours.

Results

The AM7164 differed from the newer generations across frequencies (p < 0.05) in the mechanical setup. The AM7164 produced a higher output at the lower and at the highest intensities, whereas the output was lower at the middle intensities in the mid-range compared to the newer generations. The LFE option decreased the differences at the lower frequencies, but increased differences at the higher. In free living, the mean physical activity level (PA) of the GT3X was 18 counts per minute (CPM) (8%) lower compared to the AM7164 (p < 0.001). Time spent in sedentary intensity was 26.6 minutes (95% CI 15.6 to 35.3) higher when assessed by the GT3X compared to the AM7164 (p < 0.001). Time spend in light and vigorous PA were 23.3 minutes (95% CI 31.8 to 14.8) and 11.7 minutes (95% CI 2.8 to 0.7) lower when assessed by the GT3X compared to the AM7164 (p < 0.05). When enabling the LFE the differences in the sedentary and light PA intensity (<333 counts*10 sec-1) were attenuated (p > 0.05 for differences between generations) thus attenuated the difference in mean PA (p > 0.05) when the LFE option was enabled. However, it did not attenuate the difference in time spend in vigorous PA and it introduced a difference in time spend in moderate PA (+ 3.0 min (95% CI 0.4 to 5.6)) between the generations.

Conclusion

We observed significant differences between the AM7164 and the newer Actigraph GT-generations (GT1M, GT3X and GT3X+) in a mechanical setup and in free-living. Enabling the LFE option attenuated the differences in mean PA completely, but induced a bias in the moderate PA intensities.

【 授权许可】

   
2012 Ried-Larsen et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140709072450470.pdf 2344KB PDF download
Figure 3 . 46KB Image download
Figure 2 . 46KB Image download
Figure 1 . 121KB Image download
【 图 表 】

Figure 1 .

Figure 2 .

Figure 3 .

【 参考文献 】
  • [1]Riddoch CJ, Leary SD, Ness AR, Blair SN, Deere K, Mattocks C, et al.: Prospective associations between objective measures of physical activity and fat mass in 12–14 year old children: the Avon Longitudinal Study of Parents and Children (ALSPAC). British Medical Journal 2009., 339
  • [2]Troiano RP, Berrigan D, Dodd KW, Masse LC, Tilert T, Mcdowell M: Physical activity in the United States measured by accelerometer. Medicine and Science in Sports and Exercise 2008, 40:181-188.
  • [3]Kristensen PL, Moller NC, Korsholm L, Wedderkopp N, Andersen LB, Froberg K: Tracking of objectively measured physical activity from childhood to adolescence: The European youth heart study. Scandinavian Journal of Medicine & Science in Sports 2008, 18:171-178.
  • [4]Metcalf BS, Curnow JS, Evans C, Voss LD, Wilkin TJ: Technical reliability of the CSA activity monitor: The EarlyBird Study. Med Sci Sports Exerc 2002, 34:1533-1537.
  • [5]Metcalf BS, Voss LD, Hosking J, Jeffery AN, Wilkin TJ: Physical activity at the government-recommended level and obesity-related health outcomes: a longitudinal study (early bird 37). Arch Dis Child 2008, 93:772-777.
  • [6]Rothney MP, Apker GA, Song Y, Chen KY: Comparing the performance of three generations of ActiGraph accelerometers. J Appl Physiol 2008, 105:1091-1097.
  • [7]John D, Tyo B, Bassett DR: Comparison of four ActiGraph accelerometers during walking and running. Med Sci Sports Exerc 2010, 42:368-374.
  • [8]Kozey SL, Staudenmayer JW, Troiano RP, Freedson PS: Comparison of the ActiGraph 7164 and the ActiGraph GT1M during self-paced locomotion. Med Sci Sports Exerc 2010, 42:971-976.
  • [9]Fudge BW, Wilson J, Easton C, Irwin L, Clark J, Haddow O, et al.: Estimation of oxygen uptake during fast running using accelerometry and heart rate. Med Sci Sports Exerc 2007, 39:192-198.
  • [10]Corder K, Brage S, Ramachandran A, Snehalatha C, Wareham N, Ekelund U: Comparison of two Actigraph models for assessing free-living physical activity in Indian adolescents. J Sports Sci 2007, 25:1607-1611.
  • [11]Sasaki JE, John D, Freedson PS: Validation and comparison of ActiGraph activity monitors. J Sci Med Sport 2011, 14:411-416.
  • [12]Grontved A, Hu FB: Television viewing and risk of type 2 diabetes, cardiovascular disease, and all-cause mortality: a meta-analysis. JAMA 2011, 305:2448-2455.
  • [13]Katzmarzyk PT, Church TS, Craig CL, Bouchard C: Sitting time and mortality from all causes, cardiovascular disease, and cancer. Med Sci Sports Exerc 2009, 41:998-1005.
  • [14]The Actigraph website. What is the Low Frequency Extension? 12-03-2012. http://support.theactigraph.com/faq/frequency-extension. webcite 3-12-2012. Ref Type: Online Source
  • [15]John D, Freedson P: ActiGraph and Actical Physical Activity Monitors: A Peek under the Hood. Medicine and Science in Sports and Exercise 2012, 44:S86-S89.
  • [16]Moeller NC, Korsholm L, Kristensen PL, Andersen LB, Wedderkopp N, Froberg K: Unit-specific calibration of Actigraph accelerometers in a mechanical setup - is it worth the effort? The effect on random output variation caused by technical inter-instrument variability in the laboratory and in the field. BMC Med Res Methodol 2008, 8:19. BioMed Central Full Text
  • [17]Esliger DW, Tremblay MS: Technical reliability assessment of three accelerometer models in a mechanical setup. Medicine and Science in Sports and Exercise 2006, 38:2173-2181.
  • [18]Brage S, Brage N, Froberg K, Wedderkopp N: Reliability and Validity of the Computer Science and Applications Model 7164 accelerometer in a mechanical setup. Meas Phys Edu Exerc Sci 2003, 7:101-119.
  • [19]Schepens B, Willems PA, Cavagna GA: The mechanics of running in children. Journal of Physiology-London 1998, 509:927-940.
  • [20]Kram R, Griffin TM, Donelan JM, Chang YH: Force treadmill for measuring vertical and horizontal ground reaction forces. J Appl Physiol 1998, 85:764-769.
  • [21]Matthews CE, Chen KY, Freedson PS, Buchowski MS, Beech BM, Pate RR, et al.: Amount of time spent in sedentary behaviors in the United States, 2003–2004. Am J Epidemiol 2008, 167:875-881.
  • [22]Andersen LB, Harro M, Sardinha LB, Froberg K, Ekelund U, Brage S, et al.: Physical activity and clustered cardiovascular risk in children: a cross-sectional study (The European Youth Heart Study). Lancet 2006, 368:299-304.
  • [23]Freedson PS, Melanson E, Sirard J: Calibration of the Computer Science and Applications, Inc. accelerometer. Med Sci Sports Exerc 1998, 30:777-781.
  • [24]Hendelman D, Miller K, Baggett C, Debold E, Freedson P: Validity of accelerometry for the assessment of moderate intensity physical activity in the field. Med Sci Sports Exerc 2000, 32:S442-S449.
  • [25]Yngve A, Nilsson A, Sjostrom M, Ekelund U: Effect of monitor placement and of activity setting on the MTI accelerometer output. Med Sci Sports Exerc 2003, 35:320-326.
  • [26]Bull FC, the Expert Working Groups: Physical Activity Guidelines in the U.K.: Review and Recommendations. School of Sport, Exercise and Health Sciences, Loughborough University; 2010. Ref Type: Report
  • [27]Orendurff MS, Segal AD, Klute GK, Berge JS, Rohr ES, Kadel NJ: The effect of walking speed on center of mass displacement. Journal of Rehabilitation Research and Development 2004, 41:829-834.
  • [28]Nilsson A, Ekelund U, Yngve A, Sjostrom M: Assessing physical activity among children with accelerometers using different time sampling intervals and placements. Pediatr Exerc Sci 2002, 14:87-96.
  文献评价指标  
  下载次数:52次 浏览次数:40次