Effect of lung flooding and high-intensity focused ultrasound on lung tumours: an experimental study in an ex vivo human cancer model and simulated in vivo tumours in pigs
Thomas Günther Lesser1  Sabine Bischoff2  Harald Schubert2  Carsten Boltze3  Frank Wolfram1 
[1] Department of Thoracic and Vascular Surgery, SRH Wald-Klinikum Gera, Teaching Hospital of Friedrich-Schiller-University of Jena, Strasse des Friedens 122, D-07548 Gera, Germany;Institute of Animal Experimentation, Friedrich-Schiller University of Jena, Bachstrasse 18, D-07743 Jena, Germany;Department of Pathology, SRH Wald-Klinikum Gera, Teaching Hospital of Friedrich-Schiller- University of Jena, Strasse des Friedens 122, D-07548 Gera, Germany
关键词: Lung flooding;    Lung cancer ablation;    High-intensity focused ultrasound;   
Others  :  817675
DOI  :  10.1186/2047-783X-19-1
 received in 2013-08-13, accepted in 2013-12-19,  发布年份 2014
PDF
【 摘 要 】

Background

High-intensity focused ultrasound is a valuable tool for minimally invasive tumour ablation. However, due to the air content in ventilated lungs, lung tumours have never been treated with high-intensity focused ultrasound. Lung flooding enables efficient lung sonography and tumour imaging in ex vivo human and in vivo porcine lung cancer models. The current study evaluates the effectiveness of lung flooding and sonography-guided high-intensity focused ultrasound for lung tumour ablation in ex vivo human and in vivo animal models.

Methods

Lung flooding was performed in four human lung lobes which were resected from non-small cell lung cancers. B-mode imaging and temperature measurements were simultaneously obtained during high-intensity focused ultrasonography of centrally located lung cancers. The tumour was removed immediately following insonation and processed for nicotinamide adenine dinucleotide phosphate-diaphorase and H&E staining. In addition, the left lungs of three pigs were flooded. Purified BSA in glutaraldehyde was injected centrally into the left lower lung lobe to simulate a lung tumour. The ultrasound was focused transthoracically through the flooded lung into the simulated tumour with the guidance of sonography. The temperature of the tumour was simultaneously measured. The vital signs of the animal were monitored during the procedure.

Results

A well-demarcated lesion of coagulation necrosis was produced in four of four human lung tumours. There did not appear to be any damage to the surrounding lung parenchyma. After high-intensity focused ultrasound insonation, the mean temperature increase was 7.5-fold higher in the ex vivo human tumour than in the flooded lung tissue (52.1 K ± 8.77 K versus 7.1 K ± 2.5 K). The transthoracic high-intensity focused ultrasound of simulated tumours in the in vivo model resulted in a mean peak temperature increase up to 53.7°C (±4.5). All of the animals survived the procedure without haemodynamic complications.

Conclusions

High-intensity focused ultrasound with lung flooding produced a thermal effect in an ex vivo human lung carcinoma and in vivo simulated lung tumours in a porcine model. High-intensity focused ultrasound is a potential new strategy for treating lung cancer.

【 授权许可】

   
2014 Wolfram et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140711014224293.pdf 3209KB PDF download
Figure 8. 69KB Image download
Figure 7. 176KB Image download
Figure 6. 190KB Image download
Figure 5. 123KB Image download
Figure 4. 98KB Image download
Figure 3. 44KB Image download
Figure 2. 133KB Image download
Figure 1. 46KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Figure 7.

Figure 8.

【 参考文献 】
  • [1]DKFZ: Atlas of Cancer Mortality. Heidelberg: German Center of Cancer Research; 2007.
  • [2]Drings P, Dienemann H, Wannenmacher M: Management of Lung cancer. Berlin, Heidelberg, New York: Springer; 2002.
  • [3]Colice GL, Rubins J, Unger M: Follow-up and surveillance of the lung cancer patient following curative-intent therapy. Chest 2003, 123:272-283.
  • [4]Simon CJ, Dupuy DE, DiPetrillo TA, Safran HP, Grieco CA, Ng T, Mayo-Smith WW: Pulmonary radiofrequency ablation: long-term safety and efficacy in 153 patients. Radiology 2007, 243:268-275.
  • [5]Von Meyerfeldt EM, Prevoo W, Peyrot D, Lai a Fat N, Burgers SJ, Wouters MW, Klomp HM: Local progression after radiofrequency ablation of pulmonary metastases. Cancer 2011, 117:3781-3787.
  • [6]Zhou YF: High intensity focused ultrasound in clinical tumour ablation. World J Clin Oncol 2011, 2:8-27.
  • [7]Crouzet S, Rebillard X, Chevallier D, Rischmann P, Pasticier G, Garcia G, Rouviere O, Chapelon JY, Gelet A: Multicentric oncologic outcomes of high-intensity focused ultrasound for localized prostate cancer in 803 patients. Eur Urol 2010, 58:559-566.
  • [8]Nerendra T, Sanghvi T, Rescorla F, Galliani C, Fry F, Griffith S, Grosfeld J, Yang R: Extracorporeal liver ablation using sonography-guided high intensity focused ultrasound. Investig Radiol 1992, 27:796-803.
  • [9]Crouzet S, Poissonnier L, Murat FJ, Pasticier G, Rouvière O, Mège-Lechevallier F, Chapelon JY, Martin X, Gelet A, Melodelima D, Vincenot J, Chen Y, Dupre A, Chapelon J, Rivoire M: Washington: outcomes of HIFU for localised prostate cancer using the ablatherm integrate imaging® device. Prog Urol 2011, 21:191-197.
  • [10]Mikhak Z, Pedersen P: Acoustic attenuation properties of the lung: an open question. Ultrasound Med Biol 2002, 28:1209-1216.
  • [11]Lesser TG, Schubert H, Bischoff S, Wolfram F: Lung flooding enables efficient lung sonography and tumour imaging in human ex vivo and porcine in vivo lung cancer model. Eur J Med Res 2013, 18:23. [Epub ahead of print] BioMed Central Full Text
  • [12]Snoeren N, Huiskens J, Rijken AM, van Hillegersberg R, van Erkel AR, Slooter GD, Klaase JM, van den Tol GM, Ten Kate F, Jansen MCJ, van Gulik TM: Viable tumour tissue adherent to needle applicators after local ablation: a risk factor for local tumour progression. Ann Surg Oncol 2011, 18:3702-3710.
  • [13]Kennedy J, ter Haar G, Cranston D: High intensity focused ultrasound: surgery of the future? Br J Radiol 2003, 76:590-599.
  • [14]Thüroff S, Chaussy C, Vallancien G, Wieland W, Kiel HJ, Le Duc A, Desgrandchamps F, De La Rosette JJ, Gelet A: High-intensity focused ultrasound and localized prostate cancer: efficacy results from the European multicentric study. J Endourol 2003, 17:673-677.
  • [15]Furusawa H, Namba K, Nakahara H, Tanaka C, Yasuda Y, Hirabara E, Imahariyama M, Komaki K: The evolving non-surgical ablation of breast cancer: MR guided focused ultrasound (MRgFUS). Breast Cancer 2007, 14:55-58.
  • [16]Wu F, Wang ZB, Zhu H, Chen WZ, Zou JZ, Bai J, Li KQ, Jin CB, Xie FL, Su HB: Feasibility of US-guided high-intensity focused ultrasound treatment in patients with advanced pancreatic cancer: initial experience. Radiology 2005, 236:1034-1040.
  • [17]Wu F, Wang ZB, Chen WZ, Zou JZ, Bai J, Zhu H, Li KQ, Jin CB, Xie FL, Su HB: Advanced hepatocellular carcinoma: treatment with high-intensity focused ultrasound ablation combined with transcatheter arterial embolization. Radiology 2005, 235:659-667.
  • [18]Chen W, Zhou K: High-intensity focused ultrasound ablation: a new strategy to manage primary bone tumours. Curr Opin Orthop 2005, 16:494-500.
  • [19]Treeby B, Zhang E, Thomas AS, Cox B: Measurement of the ultrasound attenuation and dispersion in whole human blood and its components from 0 to 70 MHz. Ultrasound Med Biol 2011, 37:289-300.
  • [20]Yang R, Reilly CR, Rescorla FJ, Faught PR, Sanghvi NT, Fry FJ, Franklin TD Jr, Lumeng L, Grosfeld JL: High-intensity focused ultrasound in the treatment of experimental liver cancer. Arch Surg 1991, 126:1002-1009.
  • [21]Wu CC, Chen WS, Ho MC, Huang KW, Chen CN, Yen JY, Lee PH: Minimizing abdominal wall damage during high intensity focused ultrasound ablation by inducing artificial ascites. J Acoust Soc Am 2008, 124:674-679.
  • [22]Lesser T: Neue Möglichkeiten der kompletten Lungensonographie bei einseitiger Flüssigkeitsfüllung: experimentelle Untersuchungen zur Technik, Herdsimulation, Hämodynamik und Morphologie. Jena: Universitätsbibliothek; 1999.
  • [23]Rieke V, King RL, Swaminathan A, McConnell M, Butts Pauly K: MR-guided focused ultrasound ablation through the ribcage. Intl Soc Mag Reson Med 2009, 17:2529.
  • [24]Jolesz FA, Hynynen K, McDannold N, Tempany C: MR imaging-controlled focused ultrasound ablation: a noninvasive image-guided surgery. Magn Reson Imaging Clin N Am 2005, 13:545-560.
  • [25]Ellis S, Rieke V, Kohi M, Westphalen AC: Clinical applications for magnetic resonance guided high intensity focused ultrasound (MRgHIFU): present and future. J Med Imaging Radiat Oncol 2013, 57:391-399.
  • [26]Rieke V, Butts Pauly K: MR thermometry. J Magn Reson Imaging 2008, 27:376-390.
  文献评价指标  
  下载次数:0次 浏览次数:5次