期刊论文详细信息
Journal of Cardiothoracic Surgery
The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography
Minwen Zheng2  Yuewen Hao2  Mengqi Wei2  Junqing Xu2  Hongtao Liu1  Jing Ren2  Jian Li2  Jian Xu2  Ying Liu2 
[1] Siemens China Ltd., Shang Hai, China;Department of Radiology, Xijing Hospital, Fourth Military Medical University, 169 West Changle Road, Xi’an 710032, China
关键词: Dual-source CT;    Image quality;    Radiation dose;    Aorta;   
Others  :  813918
DOI  :  10.1186/1749-8090-8-228
 received in 2013-06-27, accepted in 2013-12-09,  发布年份 2013
PDF
【 摘 要 】

Background

Aortic dissection is a lift-threatening medical emergency associated with high rates of morbidity and mortality. The incidence rate of aortic dissection is estimated at 5 to 30 per 1 million people per year. The prompt and correct diagnosis of aortic dissection is critical. This study was to compare the ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography and conventional dual-source CT angiography.

Methods

A total of 110 consecutive patients with suspected aortic dissection and other aortic disorders were randomly divided into two groups. Group A underwent traditional scan mode and Group B underwent high-pitch dual-source CT scan mode. The image quality and radiation dose of two groups were compared.

Results

Close interobserver agreement was found for image quality scores (κ = 0.87). The image quality of ascending aorta was significantly better in the high-pitch group than in the conventional group (2.78 ± 0.46 vs 1.57 ± 0.43, P < 0.001). There was no significant difference of the CT attenuation values, the aortic image noise and SNR between two groups. The mean radiation dose of high-pitch group was also significantly lower than that of conventional group (2.7 ± 0.6 mSv vs. 3.9 ± 0.9 mSv, P < 0.001).

Conclusions

High-pitch dual-source CT angiography of the whole aorta can provide motion-artifact-free imaging of the ascending aorta at a low radiation dose compared to conventional protocol.

【 授权许可】

   
2013 Liu et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140710020304206.pdf 909KB PDF download
Figure 2. 69KB Image download
Figure 1. 70KB Image download
【 图 表 】

Figure 1.

Figure 2.

【 参考文献 】
  • [1]Fujioka C, Horiguchi J, Kiguchi M, Yamamoto H, Kitagawa T, Ito K: Survey of aorta and coronary arteries with prospective ECG-triggered 100-kV 64-MDCT angiography. AJR Am J Roentgenol 2009, 193:227-233.
  • [2]Tsang W, Bateman MG, Weinert L, Pellegrini G, Mor-Avi V, Sugeng L, Yeung H, Patel AR, Hill AJ, Iaizzo PA, Lang RM: Accuracy of aortic annular measurements obtained from three-dimensional echocardiography, CT and MRI: human in vitro and in vivo studies. Heart 2012, 98:1146-1152.
  • [3]Nayeemuddin M, Pherwani AD, Asquith JR: Imaging and management of complications of open surgical repair of abdominal aortic aneurysms. Clin Radiol 2012, 67:802-814.
  • [4]Duquette AA, Jodoin PM, Bouchot O, Lalande A: 3D segmentation of abdominal aorta from CT-scan and MR images. Comput Med Imaging Graph 2012, 36:294-303.
  • [5]Khan NU, Yonan N: Does preoperative computed tomography reduce the risks associated with re-do cardiac surgery? Interact Cardiovasc Thorac Surg 2009, 9:119-123.
  • [6]Singh C, Gupta M, Sharma S: Compression of trachea due to double aortic arch: demonstration by multi-slice CT scan (MSCT). Heart Lung Circ 2006, 15:332-333.
  • [7]Willoteaux S, Lions C, Gaxotte V, Negaiwi Z, Beregi JP: Imaging of aortic dissection by helical computed tomography (CT). Eur Radiol 2004, 14:1999-2008.
  • [8]Robinson SJ, Manns RA: Thoracic aortic iIntra-mural haematoma progression to dissection demonstrated on multi-slice CT imaging: don’t forget the follow-up scan! Clin Radiol 2003, 58:739-741.
  • [9]Smith AD, Schoenhagen P: CT imaging for acute aortic syndrome. Cleve Clin J Med 2008, 75:7-9. 12, 15-17 passim
  • [10]Roos JE, Willmann JK, Weishaupt D, Lachat M, Marincek B, Hilfiker PR: Thoracic aorta: motion artifact reduction with retrospective and prospective electrocardiography-assisted multi-detector row CT. Radiology 2002, 222:271-277.
  • [11]Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF: Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging 2008, 24:535-546.
  • [12]Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, Hanson R, Khorasani R: Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology 2009, 251:175-184.
  • [13]Flohr TG, McCollough CH, Bruder H, Petersilka M, Gruber K, Süss C, Grasruck M, Stierstorfer K, Krauss B, Raupach R, Primak AN, Küttner A, Achenbach S, Becker C, Kopp A, Ohnesorge BM: First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 2006, 16:256-268.
  • [14]Matt D, Scheffel H, Leschka S, Flohr TG, Marincek B, Kaufmann PA, Alkadhi H: Dual-source CT coronary angiography: image quality, mean heart rate, and heart rate variability. AJR Am J Roentgenol 2007, 189:567-573.
  • [15]Achenbach S, Marwan M, Schepis T, Pflederer T, Bruder H, Allmendinger T, Petersilka M, Anders K, Lell M, Kuettner A, Ropers D, Daniel WG, Flohr T: High-pitch spiral acquisition: a new scan mode for coronary CT angiography. J Cardiovasc Comput Tomogr 2009, 3:117-121.
  • [16]Petersilka M, Bruder H, Krauss B, Stierstorfer K, Flohr TG: Technical principles of dual source CT. Eur J Radiol 2008, 68:362-368.
  • [17]Goetti R, Baumüller S, Feuchtner G, Stolzmann P, Karlo C, Alkadhi H, Leschka S: High-pitch dual-source CT angiography of the thoracic and abdominal aorta: is simultaneous coronary artery assessment possible? AJR Am J Roentgenol 2010, 194:938-944.
  • [18]Stinn B, Stolzmann P, Fornaro J, Hibbeln D, Alkadhi H, Wildermuth S, Leschka S: Technical principles of computed tomography in patients with congenital heart disease. Insights Imaging 2011, 2:349-356.
  • [19]Karlo C, Leschka S, Goetti RP, Feuchtner G, Desbiolles L, Stolzmann P, Plass A, Falk V, Marincek B, Alkadhi H, Baumüller S: High-pitch dual-source CT angiography of the aortic valve-aortic root complex without ECG-synchronization. Eur Radiol 2011, 21:205-212.
  • [20]Bolen MA, Popovic ZB, Tandon N, Flamm SD, Schoenhagen P, Halliburton SS: Image quality, contrast enhancement, and radiation dose of ECG-triggered high-pitch CT versus non-ECG-triggered standard-pitch CT of the thoracoabdominal aorta. AJR Am J Roentgenol 2012, 198:931-938.
  • [21]Bongartz G, Golding SJ, Jurik AG, Leonardi M, Van Persijn Van Meerten E, Rodríguez R, Schneider K, Calzado A, Geleijns J, Jessen KA, Panzer W, Shrimpton PC, Tosi G: European Guidelines for Multislice Computed Tomography. Brussels, Belgium: European Commission; 2004.
  • [22]Verhoeven EL, Paraskevas KI, Oikonomou K, Yazar O, Ritter W, Pfister K, Kasprzak P: Fenestrated and branched stent-grafts to treat post-dissection chronic aortic aneurysms after initial treatment in the acute setting. J Endovasc Ther 2012, 19:343-349.
  • [23]Svensson LG, Kouchoukos NT, Miller DC, Bavaria JE, Coselli JS, Curi MA, Eggebrecht H, Elefteriades JA, Erbel R, Gleason TG, Lytle BW, Mitchell RS, Nienaber CA, Roselli EE, Safi HJ, Shemin RJ, Sicard GA, Sundt TM III, Szeto WY, Wheatley GH III, Society of Thoracic Surgeons Endovascular Surgery Task Force: Expert consensus document on the treatment of descending thoracic aortic disease using endovascular stent-grafts. Ann Thorac Surg 2008, 85:S1-S41.
  • [24]Zipfel B, Czerny M, Funovics M, Coppi G, Ferro C, Rousseau H, Berti S, Tealdi DG, Riambau V, Mangialardi N, Sassi C, RESTORE Investigators: Endovascular treatment of patients with types A and B thoracic aortic dissection using Relay thoracic stent-grafts: results from the RESTORE patient registry. J Endovasc Ther 2011, 18:131-143.
  • [25]Tsai TT, Trimarchi S, Nienaber CA: Acute aortic dissection: perspectives from the International Registry of Acute Aortic Dissection (IRAD). Eur J Vasc Endovasc Surg 2009, 37:149-159.
  • [26]Chung JW, Park JH, Im JG, Chung MJ, Han MC, Ahn H: Spiral CT angiography of the thoracic aorta. Radiographics 1996, 16:811-824.
  • [27]Sebastià C, Pallisa E, Quiroga S, Alvarez-Castells A, Dominguez R, Evangelista A: Aortic dissection: diagnosis and follow-up with helical CT. Radiographics 1999, 19:45-60.
  • [28]Castañer E, Andreu M, Gallardo X, Mata JM, Cabezuelo MA, Pallardó Y: CT in nontraumatic acute thoracic aortic disease: typical and atypical features and complications. Radiographics 2003, 23:S93-S110.
  • [29]Beeres M, Schell B, Mastragelopoulos A, Herrmann E, Kerl JM, Gruber-Rouh T, Lee C, Siebenhandl P, Bodelle B, Zangos S, Vogl TJ, Jacobi V, Bauer RW: High-pitch dual-source CT angiography of the whole aorta without ECG synchronisation: initial experience. Eur Radiol 2012, 22(1):129-137.