BMC Research Notes
Validation of a triglyceride meter for use in pregnancy
Leonie K Callaway2  H David McIntyre1  Michael D’Emden2  Marloes Dekker Nitert2  Helen L Barrett2 
[1] Mater Medical Research Institute, Queensland, Australia;School of Medicine, The University of Queensland, St Lucia, Queensland, Australia
关键词: Point of care;    Diabetes;    Pregnancy;    Triglycerides;   
Others  :  1127396
DOI  :  10.1186/1756-0500-7-679
 received in 2014-03-03, accepted in 2014-09-15,  发布年份 2014
PDF
【 摘 要 】

Background

Elevated maternal triglycerides have been associated with adverse pregnancy outcomes including an increased risk of preeclampsia and macrosomia. A valid triglyceride meter would allow the examination of maternal postprandial triglycerides in a systematic manner. A non-fasting venous and two capillary measurements (using the Roche Accutrend® Plus meter) of triglycerides were measured in 40 participants at a mean of 36 weeks gestation.

Findings

The two methods were highly correlated (r = 0.89, P <0.0001), and the distributions were similar (mean difference 0.01 mmol/L (SD 0.47)), t = 0.18, P =0.86). Passing Bablok equation was: y = −0.01 + 0.98 × [95% CI intercept −0.51 – 0. 38; 95% CI slope 0.85-1.15). The estimated bias was −0.01 mmol/L (95% CI −0.93 – 0.91)).

Conclusions

This study demonstrated the Accutrend® Plus meter provides results that correlate strongly with the reference method, with low bias, when used in late pregnancy.

【 授权许可】

   
2014 Barrett et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150220133422185.pdf 195KB PDF download
Figure 1. 46KB Image download
【 图 表 】

Figure 1.

【 参考文献 】
  • [1]Niromanesh S, Shirazi M, Dastgerdy E, Sharbaf FR, Shirazi M, Khazaeipour Z: Association of hypertriglyceridaemia with pre-eclampsia, preterm birth, gestational diabetes and uterine artery pulsatility index. Natl Med J India 2012, 25:265-267.
  • [2]Son GH, Kwon JY, Kim YH, Park YW: Maternal serum triglycerides as predictive factors for large-for-gestational age newborns in women with gestational diabetes mellitus. Acta Obstet Gynecol Scand 2010, 89:700-704.
  • [3]Di Cianni G, Miccoli R, Volpe L, Lencioni C, Ghio A, Giovannitti MG, Cuccuru I, Pellegrini G, Chatzianagnostou K, Boldrini A, Del Prato S: Maternal triglyceride levels and newborn weight in pregnant women with normal glucose tolerance. Diabetic Med 2005, 22:21-25.
  • [4]Knopp RH, Magee MS, Walden CE, Bonet B, Benedetti TJ: Prediction of infant birth weight by GDM screening tests. Importance of plasma triglyceride. Diabetes Care 1992, 15:1605-1613.
  • [5]Jagla A, Schrezenmeir J: Postprandial triglycerides and endothelial function. Exp Clin Endocrinol Diabetes 2001, 109:S533-S547.
  • [6]Herrera E, Ortega-Senovilla H: Disturbances in lipid metabolism in diabetic pregnancy - Are these the cause of the problem? Best Pract Res Clin Endocrinol Metab 2010, 24:515-525.
  • [7]Winkler K, Wetzka B, Hoffmann MM, Friedrich I, Kinner M, Baumstark MW, Wieland H, Marz W, Zahradnik HP: Low density lipoprotein (LDL) subfractions during pregnancy: accumulation of buoyant LDL with advancing gestation. J Clin Endocrinol Metab 2000, 85:4543-4550.
  • [8]Hoffman L, Nolan C, Wilson JD, Oats JJ, Simmons D: Gestational diabetes mellitus–management guidelines. The Australasian Diabetes in Pregnancy Society. Med J Aust 1998, 169:93-97.
  • [9]Passing H, Bablok : A new biometrical procedure for testing the equality of measurements from two different analytical methods. Application of linear regression procedures for method comparison studies in clinical chemistry, Part I. J Clin Chem Clin Biochem 1983, 21:709-720.
  • [10]Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1:307-310.
  • [11]Mendez-Gonzalez J, Bonet-Marques R, Ordonez-Llanos J: Lipid profile in ambulatory subjects using 3 point of care devices and comparison with reference methods. Point of care: J Near Patient Test Technol 2010, 9:102-107.
  • [12]Scafoglieri A, Tresignie J, Provyn S, Clarys JP, Bautmans I: Reproducibility, accuracy and concordance of Accutrend Plus for measuring circulating lipid concentration in adults. Biochem Med (Zagreb) 2012, 22:100-108.
  • [13]Iovine C, Gentile A, Hattemer A, Pacioni D, Riccardi G, Rivellese AA: Self-monitoring of plasma triglyceride levels to evaluate postprandial response to different nutrients. Metabolism 2004, 53:620-623.
  • [14]Moses RG, Calvert D, Storlien LH: Evaluation of the Accutrend GCT with respect to triglyceride monitoring. Diabetes Care 1996, 19:1305-1306.
  文献评价指标  
  下载次数:18次 浏览次数:18次