期刊论文详细信息
BMC Musculoskeletal Disorders
Test-retest reliability of temporal and spatial gait characteristics measured with an instrumented walkway system (GAITRite®)
Marcus P Besser1  Cornelis JT van Uden2 
[1] Department of Physical Therapy, Thomas Jefferson University, 130 South 9th Street Suite 830, Philadelphia, PA 19107-5233, USA;Department of Physical Therapy, University Medical Center Nijmegen, P.O. Box 9101, 6500 HB Nijmegen, The Netherlands
关键词: test-retest reliability;    motion analysis;    gait;   
Others  :  1166182
DOI  :  10.1186/1471-2474-5-13
 received in 2004-02-08, accepted in 2004-05-17,  发布年份 2004
PDF
【 摘 要 】

Background

The purpose of this study was to determine the test-retest reliability of temporal and spatial gait measurements over a one-week period as measured using an instrumented walkway system (GAITRite®).

Methods

Subjects were tested on two occasions one week apart. Measurements were made at preferred and fast walking speeds using the GAITRite® system. Measurements tested included walking speed, step length, stride length, base of support, step time, stride time, swing time, stance time, single and double support times, and toe in-toe out angle.

Results

Twenty-one healthy subjects participated in this study. The group consisted of 12 men and 9 women, with an average age of 34 years (range: 19 – 59 years). At preferred walking speed, all gait measurements had ICC's of 0.92 and higher, except base of support which had an ICC of 0.80. At fast walking speed all gait measurements had ICC's above 0.89 except base of support (ICC = 0.79),

Conclusions

Spatial-temporal gait measurements demonstrate good to excellent test-retest reliability over a one-week time span.

【 授权许可】

   
2004 van Uden and Besser; licensee BioMed Central Ltd. This is an Open Access article: verbatim copying and redistribution of this article are permitted in all media for any purpose, provided this notice is preserved along with the article's original URL.

【 预 览 】
附件列表
Files Size Format View
20150416041701527.pdf 212KB PDF download
【 参考文献 】
  • [1]Krebs DE, Edelstein JE, Fishman S: Reliability of observational kinematic gait analysis. Phys Ther 1985, 65:1027-1033.
  • [2]Coutts F: Gait analysis in the therapeutic environment. Man Ther 1999, 4:2-10.
  • [3]Bilney B, Morris M, Webster K: Concurrent related validity of the GAITRite walkway system for quantification of the spatial and temporal parameters of gait. Gait Posture 2003, 17:68-74.
  • [4]Urquhart DM, Morris ME, Iansek R: Gait consistency over a 7-day interval in people with Parkinson's disease. Arch Phys Med Rehabil 1999, 80:696-701.
  • [5]Cutlip RG, Mancinelli C, Huber F, DiPasquale J: Evaluation of an instrumented walkway for measurement of the kinematic parameters of gait. Gait Posture 2000, 12:134-138.
  • [6]McDonough AL, Batavia M, Chen FC, Kwon S, Ziai J: The validity and reliability of the GAITRite system's measurements: A preliminary evaluation. Arch Phys Med Rehabil 2001, 82:419-425.
  • [7]Besser MP, Kmieczak K, Schwartz L, Snyderman M, Wasko J, Selby-Silverstein L: Reliability of temporal spatial gait parameters means in adults without imparment. In Annual Meeting of the Gait and Clinical Movement Analysis Society. Dallas, Texas, Gait Posture; 1999:9(2): 113.
  • [8]Portney LG, Watkins MP: Foundations of clinical research: applications to practice. Upper Saddle River, NJ, Prentice-Hall, Inc; 2000.
  文献评价指标  
  下载次数:6次 浏览次数:16次