Arthroscopic suture retrievers and shuttles: a biomechanical investigation of the force required for tendon penetration and defect size
Dominik C. Meyer1  Georg Lajtai2  Karl Wieser1  Christopher G. Lenz1 
[1]Orthopaedic Department, Balgrist Hospital, University of Zurich, Forchstrasse 340, Zurich, CH-8008, Switzerland
[2]Orthopaedic Department, Private Hospital Maria Hilf, Klagenfurt, Austria
关键词: Iatrogenic damage;    Lesion size;    Suture shuttle;    Suture retriever;    Tendon;    Rotator cuff;   
Others  :  1233988
DOI  :  10.1186/s12891-015-0794-9
 received in 2014-12-10, accepted in 2015-10-29,  发布年份 2015
PDF
【 摘 要 】

Background

To compare instruments designed for arthroscopic suture handling during arthroscopic rotator cuff repair, to assess the force needed to penetrate the tendon, and to evaluate the residual defect size.

Methods

Twenty-one instruments were each tested ten times on thawed sheep infraspinatus tendons. The force needed to pierce the tendon with each instrument was measured using a custom setup. Bone wax plates were used to make the perforation marks visible and to quantify the lesions each instrument created.

Results

The force to pierce a tendon had a range of 5.6–18.5 N/mm. Within the group of suture retrievers, the angled instruments required in average 85 % higher forces than straight instruments. The lesion area had a range of 2–7 mm 2 . Suture retrievers produced significantly larger lesion sizes compared with suture shuttles.

Conclusion

For the identical task of passing a suture through a tendon, differences exist regarding the ease of tendon penetration and potential damage to the tendon for different tools. The design, function, and resulting lesion size may be relevant and important for surgical handling and to avoid excess structural damage to the tendon. These results suggest that choosing the most appropriate tools for arthroscopic suture stitching influences the ease of handling and final integrity of the tissue.

Level of evidence

Mechanical evaluation of surgical devices.

【 授权许可】

   
2015 Lenz et al.

【 预 览 】
附件列表
Files Size Format View
20151125023904609.pdf 2551KB PDF download
Fig. 5. 48KB Image download
Fig. 4. 44KB Image download
Fig. 3. 36KB Image download
Fig. 2. 27KB Image download
Fig. 1. 36KB Image download
【 图 表 】

Fig. 1.

Fig. 2.

Fig. 3.

Fig. 4.

Fig. 5.

【 参考文献 】
  • [1]Ozbaydar M, Chung S, Diller D, Warner JJ. Arthroscopic reconstruction of the rotator cuff. The current gold standard? Orthopade. 2007; 36(9):825-33.
  • [2]Via AG, De Cupis M, Spoliti M, Oliva F. Clinical and biological aspects of rotator cuff tears. Muscles Ligaments Tendons J. 2013; 3(2):70-9.
  • [3]Murray TFJ, Lajtai G, Mileski RM. Arthroscopic repair of medium to large full-thickness rotator cuff tears: Outcome at 2- to 6-year follow-up. J Shoulder Elbow Surg. 2002; 11:19-24.
  • [4]Cho NS, Yi JW, Lee BG, Rhee YG. Retear patterns after arthroscopic rotator cuff repair: single-row versus suture bridge technique. Am J Sports Med. 2010; 38(4):664-71.
  • [5]Cho NS, Lee BG, Rhee YG. Arthroscopic rotator cuff repair using a suture bridge technique: is the repair integrity actually maintained? Am J Sports Med. 2011; 39(10):2108-16.
  • [6]Rawson S, Cartmell S, Wong J. Suture techniques for tendon repair; a comparative review. Muscles Ligaments Tendons J. 2013; 3(3):220-8.
  • [7]Chokshi BV, Kubiak EN, Jazrawi LM, Ticker JB, Zheng N, Kummer FJ, Rokito AS. The effect of arthroscopic suture passing instruments on rotator cuff damage and repair strength. Bull Hosp Jt Dis. 2006; 63(3–4):123-5.
  • [8]Cummins CA, Murrell GAC. Mode of failure for rotator cuff repair with suture anchors identified at revision surgery. J Shoulder Elbow Surg. 2003; 12(2):128-33.
  • [9]Wieser K, Rahm S, Farshad M, Ek ET, Gerber C, Meyer DC. Stitch positioning influences the suture hold in supraspinatus tendon repair. Knee Surg Sports Traumatol Arthrosc. 2013; 21(7):1587-92.
  • [10]Trantalis JN, Boorman RS, Pletsch K, Lo IK. Medial rotator cuff failure after arthroscopic double-row rotator cuff repair. Arthroscopy. 2008; 24:727-31.
  文献评价指标  
  下载次数:36次 浏览次数:30次