BMC Research Notes
A validation study: assessing the reliability of the hand held StatStripXPress lactate meter to test lactate in amniotic fluid
Sally K Tracy3  Mark B Tracy4  William D Rawlinson1  Diana D Wong1  Beverley Hall2 
[1] School of Medical Sciences, Faculty of Medicine, University of New South Wales, Kensington, NSW 2031, Australia;Department of Clinical Chemistry, South Eastern Area Laboratory Services, Randwick 2031, Australia;Midwifery and Women’s Health Research Unit, Royal Hospital for Women, 1st Floor, Randwick, NSW 2031, Australia;Westmead NICU, Westmead Hospital, Westmead, NSW 2145, Australia
关键词: Point of care;    Labour;    Amniotic fluid;    Lactate;    Dystocia;   
Others  :  1091362
DOI  :  10.1186/1756-0500-7-935
 received in 2014-08-12, accepted in 2014-12-12,  发布年份 2014
PDF
【 摘 要 】

Background

The level of lactate in amniotic fluid may provide useful clinical information when assessing whether a woman in labour is experiencing labour dystocia. If so, a rapid, reliable method to assess the concentration of amniotic fluid lactate at the bedside will be required in order to be clinically relevant. To assess efficacy, we compared the hand held StatStripXPreass lactate meter (Nova Biomedical) to the reference laboratory analyser ABX Pentra 400 (Horiba) in a controlled environment. Baseline biological lactate concentration was measured in triplicate and samples of a known quantity of thawed amniotic fluid spiked with lactate substrate (62 mmol/L) from the LDH12 kit (Roche, SUI) to yield a predetermined lactate concentration above baseline then measured in triplicate. Deming Regression was used to determine the linear agreement and a Bland Altman plot used to determine the paired agreement across the range of values.

Findings

The mean difference with Bland-Altman plot between hand held meter and lab instrument was -1.0 mmol/L (SD 3.0 mmol/L) with 95% CI limits of agreement between -6.9 mmol/L to 4.9 mmol/L. The Deming regression co-efficient or slope of agreement was 0.91 (SD of 0.21).

Conclusion

The measurement of amniotic fluid lactate using the StatStripXPress hand held meter was reliable compared to reference laboratory methods for measuring lactate levels in amniotic fluid.

【 授权许可】

   
2014 Hall et al.; licensee BioMed Central.

【 预 览 】
附件列表
Files Size Format View
20150128171422829.pdf 210KB PDF download
Figure 2. 22KB Image download
Figure 1. 32KB Image download
【 图 表 】

Figure 1.

Figure 2.

【 参考文献 】
  • [1]Lowe N: A review of factors associated with dystocia and caserean section in nulliparous women. J Midwifery Womens Health 2007, 52:216-228.
  • [2]Li Z, McNally L, Hilder L, Sullivan E: Australia’s mothers and babies. In Perinatal statistics series. Sydney: Australian Institute of Health and Werlfare; 2009.
  • [3]Kjaergaard H, Olsen J, Ottesen B, Nyberg P, Dykes A-K: Obstetric risk indicators for labour dystocia in nulliparous women: a multi-centre cohort study. BMC Pregnancy Childbirth 2008, 8(1):45. BioMed Central Full Text
  • [4]Wiberg-Itzel E, Pettersson H, Cnattingius S, Nordstrom L: Association between lactate concentration in amniotic fluid and dysfunctional labor. Acta Obstet Gynecol Scand 2008, 87:924-928.
  • [5]Wiberg-Itzel E, Pettersson H, Andolf E, Hansson A, Winbladh B, Akerud H: Lactate concentration in amniotic fluid: a good predictor of labor outcome. Eur J Obstet Gynecol Reprod Biol 2010, 152:34-38.
  • [6]Wiberg-Itzel E, Akerud H, Andolf E, Hellstrom-Westas L, Winbladh B, Wennerholm UB: Association between adverse neonatal outcome and lactate concentration in amniotic fluid. Obstet Gynecol 2011, 118:135-142.
  • [7]Hall B, Iwasenko J, Moriatis M, Rawlinson W, Tracy M, Tracy SK: A pilot study to determine the feasibility of collecting amniotic fluid samples from women during labour and measuring amniotic fluid lactate at point of care. BMC Res Notes 2013, 6:112. BioMed Central Full Text
  • [8]Wiberg-Itzel E, Cnattingius S, Nordstrom L: Lactate determination in vaginal fluids: a new method in the diagnosis of prelabour rupture of membranes. Br J Gynaecol 2005, 112:754-758.
  • [9]Martin R: General deming regression for estimating systematic bias and its confidence interval in method-comparison studies. Clin Chem 2000, 46:100-104.
  • [10]Bland MJ, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 327:307-310.
  • [11]Nordstrom L: Lactate measurements in scalp and arterial cord blood. Obstet Gynaecol 2001, 13:141-145.
  • [12]Pennell CE, Tracy MB, Donald MC: Accurate lactate-A new method for rapid pont of sample measurement of lactate in fetal and neonatal blood. Neonatal Intensive Care 2000, 13:53-58.
  文献评价指标  
  下载次数:14次 浏览次数:20次