| BMC Nephrology | |
| Exsanguination of a home hemodialysis patient as a result of misconnected blood-lines during the wash back procedure: A case report | |
| Mark R Marshall2  Christopher J Hood2  Balaji Jagannathan1  Kerryanne Allcock2  | |
| [1] Renal Service, North Shore Hospital, Waitemata District Health Board, Private Bag 93503, Takapuna, Auckland, 0740, New Zealand;Department of Renal Medicine, Middlemore Hospital, Counties Manukau District Health Board, Private Bag 93311, Otahuhu, Auckland, 1640, New Zealand | |
| 关键词: Complication; Exsanguination; Home hemodialysis; | |
| Others : 1083185 DOI : 10.1186/1471-2369-13-28 |
|
| received in 2011-12-12, accepted in 2012-05-03, 发布年份 2012 | |
PDF
|
|
【 摘 要 】
Background
Home hemodialysis is common in New Zealand and associated with lower cost, improved survival and better patient experience. We present the case of a fully trained home hemodialysis patient who exsanguinated at home as a result of an incorrect wash back procedure.
Case presentation
The case involves a 67 year old male with a history of well controlled hypertension and impaired glucose tolerance. He commenced on peritoneal dialysis in 2006 following the development of end stage kidney failure secondary to focal segmental glomerulosclerosis. He transferred to hemodialysis due to peritoneal membrane failure in 2010, and successfully trained for home hemodialysis over a 20 week period. Following one month of uncomplicated dialysis at home, he was found deceased on his machine at home in the midst of dialysis. His death occurred during the wash back procedure performed using the “open circuit” method, and resulted from misconnection of the saline bag to the venous end of the extracorporeal blood circuit instead of the arterial end. This led to approximately 2.3L of his blood being pumped into the saline bag resulting in hypovolaemic shock and death from exsanguination.
Conclusions
Despite successful training, critical procedural errors can still be made by patients on home hemodialysis. In this case, the error involved misconnection of the saline bag for wash back. This case should prompt providers of home hemodialysis to review their training protocols and manuals. Manufacturers of dialysis machinery should be encouraged to design machines specifically for home hemodialysis, and consider distinguishing the arterial and venous ends of the extracorporeal blood circuit with colour coding or incompatible connectivity, to prevent occurrences such as these in the future.
【 授权许可】
2012 Allcock et al.; licensee BioMed Central Ltd.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 20150101011633684.pdf | 1134KB | ||
| Figure 2. | 25KB | Image | |
| Figure 4. | 52KB | Image | |
| Figure 3. | 66KB | Image | |
| Figure 2. | 68KB | Image | |
| Figure 1. | 67KB | Image |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
Figure 2.
【 参考文献 】
- [1]Ashton T, Marshall MR: The organization and financing of dialysis and kidney transplantation services in New Zealand. Int J Health Care Finance Econ 2007, 7:233-252.
- [2]Blagg CR: Home hemodialysis: 'home, home, sweet, sweet home!'. Nephrology (Carlton) 2005, 10:206-214.
- [3]Lynn KL, Buttimore AL: Future of home hemodialysis in Australia and New Zealand. Nephrology (Carlton) 2005, 10:231-233.
- [4]Marshall MR, Hawley CM, Kerr PG, Polkinghorne KR, Marshall RJ, Agar JW, McDonald SP: Home hemodialysis and mortality risk in Australian and New Zealand populations. Am J Kidney Dis 2011, 58:782-793.
- [5]McGregor DO, Buttimore AL, Lynn KL: Home hemodialysis: excellent survival at less cost, but still underutilized. Kidney Int 2000, 57:2654-2655.
- [6]Woods JD, Port FK, Stannard D, Blagg CR, Held PJ: Comparison of mortality with home hemodialysis and center hemodialysis: a national study. Kidney Int 1996, 49:1464-1470.
- [7]Hawley CM, Jeffries J, Nearhos J, Van Eps C: Complications of home hemodialysis. Hemodial Int 2008, 12(Suppl 1):S21-S25.