BMC Infectious Diseases
Determination of decimal reduction time (D value) of chemical agents used in hospitals for disinfection purposes
Alzira M da S Martins1  Thereza Christina Vessoni Penna1  Priscila Gava Mazzola1 
[1] Department of Biochemical and Pharmaceutical Technology, School of Pharmaceutical Sciences, University of São Paulo, São Paulo, Brazil
关键词: Bacillus stearothermophilus;    Bacillus subtilis;    Staphylococcus aureus;    Escherichia coli;    decimal reduction time (D-value);    Disinfectants;   
Others  :  1177114
DOI  :  10.1186/1471-2334-3-24
 received in 2003-02-05, accepted in 2003-10-17,  发布年份 2003
PDF
【 摘 要 】

Background

Prior to the selection of disinfectants for low, intermediate and high (sterilizing) levels, the decimal reduction time, D-value, for the most common and persistent bacteria identified at a health care facility should be determined.

Methods

The D-value was determined by inoculating 100 mL of disinfecting solution with 1 mL of a bacterial suspension (104 – 105 CFU/mL for vegetative and spore forms). At regular intervals, 1 mL aliquots of this mixture were transferred to 8 mL of growth media containing a neutralizing agent, and incubated at optimal conditions for the microorganism.

Results

The highest D-values for various bacteria were determined for the following solutions: (i) 0.1% sodium dichloroisocyanurate (pH 7.0) – E. coli and A. calcoaceticus (D = 5.9 min); (ii) sodium hypochlorite (pH 7.0) at 0.025% for B. stearothermophilus (D = 24 min), E. coli and E. cloacae (D = 7.5 min); at 0.05% for B. stearothermophilus (D = 9.4 min) and E. coli (D = 6.1 min) and 0.1% for B. stearothermophilus (D = 3.5 min) and B. subtilis (D = 3.2 min); (iii) 2.0% glutaraldehyde (pH 7.4) – B. stearothermophilus, B. subtilis (D = 25 min) and E. coli (D = 7.1 min); (iv) 0.5% formaldehyde (pH 6.5) – B. subtilis (D = 11.8 min), B. stearothermophilus (D = 10.9 min) and A. calcoaceticus (D = 5.2 min); (v) 2.0% chlorhexidine (pH 6.2) – B. stearothermophilus (D = 9.1 min), and at 0.4% for E. cloacae (D = 8.3 min); (vi) 1.0% Minncare® (peracetic acid and hydrogen peroxide, pH 2.3) – B. stearothermophilus (D = 9.1 min) and E. coli (D = 6.7 min).

Conclusions

The suspension studies were an indication of the disinfectant efficacy on a surface. The data in this study reflect the formulations used and may vary from product to product. The expected effectiveness from the studied formulations showed that the tested agents can be recommended for surface disinfection as stated in present guidelines and emphasizes the importance and need to develop routine and novel programs to evaluate product utility.

【 授权许可】

   
2003 Mazzola et al; licensee BioMed Central Ltd. This is an Open Access article: verbatim copying and redistribution of this article are permitted in all media for any purpose, provided this notice is preserved along with the article's original URL.

【 预 览 】
附件列表
Files Size Format View
20150428102700496.pdf 298KB PDF download
Figure 1. 12KB Image download
【 图 表 】

Figure 1.

【 参考文献 】
  • [1]Rutala WA, et al.: Principles of Disinfecting Patient-Care Items. In Disinfection, Sterilization and Antiseptics in Health Care. Edited by William A Champlain. Rutala: Polyscience Publishers; 1998:133-149.
  • [2]Jacobs PT, Wang JH, Gorham RA, Roberts CG: In: Disinfection, Sterilization and Antiseptics in Health Care. Champlain. Edited by William A. Rutala: Polyscience Publishers; 1998:165-166.
  • [3]Madigan MT, Martinko JM, Parker J: Brock Biology of Microorganisms. 10th edition. Pearson Education Publishers; 2002:696-707.
  • [4]Brazilian Ministry of Health Regulatory Agency – ANVISA-Agência Nacional de Vigilância Sanitária [http://www.anvisa.gov.br/] webcite 2003.
  • [5]National Air Duct Cleaners Association[http://www.nadca.com/] webcite
  • [6]Regulatory and Environmental Consulting Choosing The Proper Sanitizer Or Disinfectant [http://www.schiff-consulting.com/choosing.html] webcite
  • [7]ISO 14937 – International Standard: Sterilization of Health Care Products – General Requirements for Characterization of a Sterilizing Agent and the Development, Validation and Routine Control of a Sterilization Process for Medical Devices. 2000.
  • [8]ISO 11134 – International Standard: Sterilization of Health Care Products – Requirements for Validation and Routine Control – Industrial Moist Heat Sterilization. 1994.
  • [9]Spaulding EH: Chemical disinfection of medical and surgical materials. In Disinfection, sterilization and preservation. Philadelphia: Lea & Febiger; 1968:517-31.
  • [10]Croshaw B: Disinfectant Testing – with Particular Reference to the Rideal-Walker and Kelsey-Sykes Tests. In In Disinfectants. Their use and evaluation of effectiveness. Edited by Collins CH. New York Academic Press; 1981:1-14.
  • [11]The Brazilian Ministry of Health Directives n° 15/MS/ANVS n° 211/MS/SNVS 1999, n° 196/MS/AVNS 1983, n° 113/MS/AVNS 1993, n° 122/DTN 1993, n° 211/MS/ANVS 1999, Health Sanitation Code n°12342, 1978, Decree n° 930 of the 27th of August 1992. Disposition concerning the names for the Control of Hospital Infections Official Newspaper of the Union; Brazillian Section 112279-81. 4th September 1992
  • [12]News Database [http://www.estadao.com.br/agestado/] webcite
  • [13]Brazilian Pharmacy Council: PAHO: Program to reduce deaths of babies in Brazil. Pharmacia Brasileira 2000, 3:34.
  • [14]Penna TC, Mazzola PG, Martins AM: The efficacy of chemical agents in cleaning and disinfection programs. BMC Infect Dis 2001, 1:16. BioMed Central Full Text
  • [15]Penna TCV, Machoshvili IA, Taqueda MES: Bacillus stearothermophilus sporulation response to different composition media. PDA J Pharm Sci and Techn 1998, 52(5):198-208.
  • [16]Penna TCV, Schaffnern D, Abe LE, Machoshvili IA: Inactivation of Brazilian wild type and enterotoxigenic Escherichia coli by chlorine. J Ind Microbiol 1996, 16:57-61.
  • [17]Russell AD: Neutralization Procedures in the Evaluation of Bacterial Activity. In Desinfectants. Their use and Evaluation of Effectiveness. Edited by Collins CH. New York Academic Press; 1981:45-57.
  • [18]Graham GS, Boris CA: Chemical and Biologica Indicators. In Sterilization Technology, A Pratical Guide for Manufacturers and Users of Health Care Products. Edited by Robert F Morrisey and G Briggs Phillipis. New York: Van Nostrand Reinhold; 1993:36-70.
  • [19]Rutala WA: APIC Guideline for selection and use of disinfectants. AJIC Am J Infect Control 1995, 23(30):35A-65A.
  • [20]Gruendeman BJ, Larson EL: Antiseptics in Current Practice. In Disinfection, Sterilization and Antiseptics in Health Care. Edited by William A Champlain. Rutala: Polyscience Publishers; 1998:183-195.
  • [21]APIC guide line for handwashing and hand antisepsis in health care settings Am J Infect Control 1995, 23:251-269.
  • [22]Penna TCV, Martins AM, Mazzola PG: Identification of bacteria in drinking and purified water during the monitoring of a typical water purification system. J BMC Public Health 2002, 2:13. BioMed Central Full Text
  • [23]Rutala WA, Weber DJ: FDA labeling requirements for disinfection of endoscopes: a counterpoint. Infect Control Hosp Epidemiol 1995, 16:231-235.
  • [24]Bond WW: Endoscopes reprocessing. In Disinfection, Sterilization and Antiseptics in Health Care. Edited by William A Champlain. Rutala: Polyscience Publishers; 1998:159.
  • [25]McDonnell G, Russell AD: Antiseptics and disinfectants: activity, action, and resistance. Clin Microbiology Review 1998, 12(I):147-179.
  • [26]Penna TCV, et al.: Cleaning of Blood-Contaminated Reprocessed Angiographic Catheters and Spinal Needles. Infect Control Hosp Epidemiol 2000, 21:499-504.
  • [27]Warren IC, Hutchinson M, Ridgway JW: Comparative Assment of Swimming Pool Disinfectants. In Desinfectants. Their use and Evaluation of Effectiveness. Edited by CH Collins. New York Academic Press; 1981:45-57.
  文献评价指标  
  下载次数:49次 浏览次数:18次