BMC Musculoskeletal Disorders
Reinforcing the role of the conventional C-arm - a novel method for simplified distal interlocking
R Geoff Richards3  Meir Liebergall1  Benno Dicht2  Ladina Fliri2  Josh Schroeder1  Markus Windolf2 
[1] Orthopedic department Hadassah Hebrew University Medical Center, Jerusalem, Israel;AO Research Institute Davos, Clavadelerstrasse 8, 7270 Davos, Switzerland;Cardiff School of Biosciences, Cardiff University, Wales, UK
关键词: Computer aided surgery;    Free-hand locking;    Nailing;    Distal targeting;    Distal interlocking;   
Others  :  1150720
DOI  :  10.1186/1471-2474-13-8
 received in 2011-08-10, accepted in 2012-01-25,  发布年份 2012
PDF
【 摘 要 】

Background

The common practice for insertion of distal locking screws of intramedullary nails is a freehand technique under fluoroscopic control. The process is technically demanding, time-consuming and afflicted to considerable radiation exposure of the patient and the surgical personnel. A new concept is introduced utilizing information from within conventional radiographic images to help accurately guide the surgeon to place the interlocking bolt into the interlocking hole. The newly developed technique was compared to conventional freehand in an operating room (OR) like setting on human cadaveric lower legs in terms of operating time and radiation exposure.

Methods

The proposed concept (guided freehand), generally based on the freehand gold standard, additionally guides the surgeon by means of visible landmarks projected into the C-arm image. A computer program plans the correct drilling trajectory by processing the lens-shaped hole projections of the interlocking holes from a single image. Holes can be drilled by visually aligning the drill to the planned trajectory. Besides a conventional C-arm, no additional tracking or navigation equipment is required.

Ten fresh frozen human below-knee specimens were instrumented with an Expert Tibial Nail (Synthes GmbH, Switzerland). The implants were distally locked by performing the newly proposed technique as well as the conventional freehand technique on each specimen. An orthopedic resident surgeon inserted four distal screws per procedure. Operating time, number of images and radiation time were recorded and statistically compared between interlocking techniques using non-parametric tests.

Results

A 58% reduction in number of taken images per screw was found for the guided freehand technique (7.4 ± 3.4) (mean ± SD) compared to the freehand technique (17.6 ± 10.3) (p < 0.001). Total radiation time (all 4 screws) was 55% lower for the guided freehand technique compared to conventional freehand (p = 0.001). Operating time per screw (from first shot to screw tightened) was on average 22% reduced by guided freehand (p = 0.018).

Conclusions

In an experimental setting, the newly developed guided freehand technique for distal interlocking has proven to markedly reduce radiation exposure when compared to the conventional freehand technique. The method utilizes established clinical workflows and does not require cost intensive add-on devices or extensive training. The underlying principle carries potential to assist implant positioning in numerous other applications within orthopedics and trauma from screw insertions to placement of plates, nails or prostheses.

【 授权许可】

   
2012 Windolf et al; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150405214607296.pdf 1464KB PDF download
Figure 4. 47KB Image download
Figure 3. 40KB Image download
Figure 2. 30KB Image download
Figure 1. 216KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

【 参考文献 】
  • [1]Lam SW, Teraa M, Leenen LP, van der Heijden GJ: Systematic review shows lowered risk of nonunion after reamed nailing in patients with closed tibial shaft fractures. Injury 2010, 41:671-675.
  • [2]Kirousis G, Delis H, Megas P, Lambiris E, Panayiotakis G: Dosimetry during intramedullary nailing of the tibia. Acta Orthop 2009, 80:568-572.
  • [3]Gugala Z, Nana A, Lindsey RW: Tibial intramedullary nail distal interlocking screw placement: comparison of the free-hand versus distally-based targeting device techniques. Injury 2001, 32:21-25.
  • [4]Goulet JA, Londy F, Saltzman CL, Matthews LS: Interlocking intramedullary nails. An improved method of screw placement combining image intensification and laser light. Clin Orthop Relat Res 1992, 281:199-203.
  • [5]Krettek C, Konemann B, Miclau T, Kolbli R, Machreich T, Tscherne H: A mechanical distal aiming device for distal locking in femoral nails. Clin Orthop Relat Res 1999, 364:267-275.
  • [6]Anastopoulos G, Ntagiopoulos PG, Chissas D, Papaeliou A, Asimakopoulos A: Distal locking of tibial nails: a new device to reduce radiation exposure. Clin Orthop Relat Res 2008, 466:216-220.
  • [7]Zheng G, Zhang X, Haschtmann D, Gedet P, Langlotz F, Nolte LP: Accurate and reliable pose recovery of distal locking holes in computer-assisted intra-medullary nailing of femoral shaft fractures: a preliminary study. Comput Aided Surg 2007, 12:138-151.
  • [8]Khoury A, Liebergall M, Weil Y, Mosheiff R: Computerized fluoroscopic-based navigation-assisted intramedullary nailing. Am J Orthop (Belle Mead NJ) 2007, 36:582-585.
  • [9]Suhm N, Messmer P, Zuna I, Jacob LA, Regazzoni P: Fluoroscopic guidance versus surgical navigation for distal locking of intramedullary implants. A prospective, controlled clinical study. Injury 2004, 35:567-574.
  • [10]Gautheron T, Leitner F, Gautheron C, Ernotte D: Navigation with ultra-sound for intra-medullary nailing. In 2005. Berlin, Germany; 2005:680-683.
  • [11]Vandeweerd JM, Perrin R, Launois T, Brogniez L, Clegg PD, Desbrosse FG: Use of computed tomography in standing position to identify guidelines for screw insertion in the distal phalanx of horses: an ex vivo study. Vet Surg 2009, 38:373-379.
  • [12]Tornetta P, Patel P, Ricci WM: Distal Locking Using an Electromagnetic Field Guided Computer Based Real Time System. San Diego, California; 2009. October, 2009
  • [13]Rivkin G, Liebergall M: Challenges of technology integration and computer-assisted surgery. J Bone Joint Surg Am 2009, 91:13-16.
  • [14]Park K, Kim K, Choi YS: Comparison of mechanical rigidity between plate augmentation leaving the nail in situ and interlocking nail using cadaveric fracture model of the femur. Int Orthop 2010, 35:581-585.
  • [15]Barry TP: Radiation exposure to an orthopedic surgeon. Clin Orthop Relat Res 1984, 182:160-164.
  • [16]Singer G: Occupational radiation exposure to the surgeon. J Am Acad Orthop Surg 2005, 13:69-76.
  • [17]Chou LB, Cox CA, Tung JJ, Harris AH, Brooks-Terrell D, Sieh W: Prevalence of cancer in female orthopaedic surgeons in the United States. J Bone Joint Surg Am 2010, 92:240-244.
  • [18]International Commision on Radiological Protection (ICRP): The 2007 Recommendations of the International Commision on Radiological Protection. ICRP Publication103 2007.
  • [19]Kendoff D, Citak M, Gaulke R, Gardner MJ, Geerling J, Krettek C, et al.: Navigation for placement of scaphoid screws: a new indication for intraoperative 3D navigation-a cadaver study. Unfallchirurg 2007, 110:745-750.
  • [20]Wang JQ, Zhao CP, Wang MY, Su YG, Hu L, Sun L, et al.: Computer-assisted auto-frame navigation system for distal locking of tibial intramedullary nails: a preliminary report on clinical application. Chin J Traumatol 2006, 9:138-145.
  • [21]Holistein S, Ruth JT: Electromagnetic Distal Targeting for the Placement of Transverse Interlocking Screws. Boston, MA; 1996. September, 1996
  • [22]Feldkamp LA, Davis LC, Kress JW: Practical cone-beam algorithm. J Opt Soc Am A 1984, 1:612-619.
  • [23]Junejo F, Bouazza-Marouf K, Kerr D, Taylor AJ, Taylor GJ: X-ray-based machine vision system for distal locking of intramedullary nails. Proc Inst Mech Eng [H] 2007, 221:365-375.
  • [24]Leloup T, El Kazzi W, Schuind F, Warzee N: A novel technique for distal locking of intramedullary nail based on two non-constrained fluoroscopic images and navigation. IEEE Trans Med Imaging 2008, 27:1202-1212.
  文献评价指标  
  下载次数:349次 浏览次数:41次