| Use of UV-C radiation to disinfect non-critical patient care items: a laboratory assessment of the Nanoclave Cabinet | |
| A Peter R Wilson3  Adam P Fraise4  John C Hartley1  Martyn AC Wilkinson4  Christina R Bradley4  Elaine A Cloutman-Green1  Shanom Ali2  Ginny Moore2  | |
| [1] Great Ormond Street Hospital for Sick Children, London, UK;Department of Microbiology, UCLH Environmental Research Group, Royal Free Hampstead NHS Trust, London, NW3 2QG, UK;Clinical Microbiology and Virology, University College London Hospitals NHS Foundation Trust, London, UK;Hospital Infection Research Laboratory, Queen Elizabeth Hospital, Birmingham, UK | |
| 关键词: Adenovirus; Nosocomial pathogens; Surface disinfection; Ultraviolet radiation; | |
| Others : 1175317 DOI : 10.1186/1471-2334-12-174 |
|
| received in 2012-03-26, accepted in 2012-07-25, 发布年份 2012 | |
PDF
|
|
【 摘 要 】
Background
The near-patient environment is often heavily contaminated, yet the decontamination of near-patient surfaces and equipment is often poor. The Nanoclave Cabinet produces large amounts of ultraviolet-C (UV-C) radiation (53 W/m2) and is designed to rapidly disinfect individual items of clinical equipment. Controlled laboratory studies were conducted to assess its ability to eradicate a range of potential pathogens including Clostridium difficile spores and Adenovirus from different types of surface.
Methods
Each test surface was inoculated with known levels of vegetative bacteria (106 cfu/cm2), C. difficile spores (102-106 cfu/cm2) or Adenovirus (109 viral genomes), placed in the Nanoclave Cabinet and exposed for up to 6 minutes to the UV-C light source. Survival of bacterial contaminants was determined via conventional cultivation techniques. Degradation of viral DNA was determined via PCR. Results were compared to the number of colonies or level of DNA recovered from non-exposed control surfaces. Experiments were repeated to incorporate organic soils and to compare the efficacy of the Nanoclave Cabinet to that of antimicrobial wipes.
Results
After exposing 8 common non-critical patient care items to two 30-second UV-C irradiation cycles, bacterial numbers on 40 of 51 target sites were consistently reduced to below detectable levels (≥ 4.7 log10 reduction). Bacterial load was reduced but still persisted on other sites. Objects that proved difficult to disinfect using the Nanoclave Cabinet (e.g. blood pressure cuff) were also difficult to disinfect using antimicrobial wipes. The efficacy of the Nanoclave Cabinet was not affected by the presence of organic soils. Clostridium difficile spores were more resistant to UV-C irradiation than vegetative bacteria. However, two 60-second irradiation cycles were sufficient to reduce the number of surface-associated spores from 103 cfu/cm2 to below detectable levels. A 3 log10 reduction in detectable Adenovirus DNA was achieved within 3 minutes; after 6 minutes, viral DNA was undetectable.
Conclusion
The results of this study suggest that the Nanoclave Cabinet can provide rapid and effective disinfection of some patient-related equipment. However, laboratory studies do not necessarily replicate ‘in-use’ conditions and further tests are required to assess the usability, acceptability and relative performance of the Nanoclave Cabinet when used in situ.
【 授权许可】
2012 Moore et al.; licensee BioMed Central Ltd.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| 20150428024652492.pdf | 648KB | ||
| Figure 4. | 80KB | Image | |
| Figure 3. | 45KB | Image | |
| Figure 2. | 41KB | Image | |
| Figure 1. | 77KB | Image |
【 图 表 】
Figure 1.
Figure 2.
Figure 3.
Figure 4.
【 参考文献 】
- [1]Sexton T, Clarke P, O’Neill EO, Dillane T, Humphreys H: Environmental reservoirs of methicillin-resistant Staphylococcus aureus in isolation rooms: correlation with patient isolates and implications for hospital hygiene. J Hosp Infect 2006, 62:187-194.
- [2]Hayden MK, Bonten MJM, Blom DW, Lyle EA, van de Vijver DAMC, Weinstein RA: Reduction in acquisition of vancomycin-resistant enterococcus after enforcement of routine environmental cleaning measures. Clin Infect Dis 2006, 42:1552-1560.
- [3]Manian FA, Meyer L, Jenne J: Clostridium difficile contamination of blood pressure cuffs: a call for a closer look at gloving practices in the era of universal precautions. Infect Control Hosp Epidemiol 1996, 17:180-182.
- [4]Rutala WA: Disinfection and sterilization of patient-care items. Infect Control Hosp Epidemiol 1996, 17:377-384.
- [5]Gould D: The fall and rise of cleanliness in British healthcare and the nursing contribution. J Res Nurs 2005, 10:495-509.
- [6]Walker N, Gupta R, Cheesbrough J: Blood pressure cuffs: friend or foe? J Hosp Infect 2006, 63:167-169.
- [7]Hijnen WAM, Beerendonk EF, Medema GJ: Inactivation credit of UV radiation for viruses, bacteria and protozoan (oo)cysts in water: A review. Wat Res 2006, 40:3-22.
- [8]Andersen BM, Bånrud H, Bøe E, Bjordal O, Drangsholt F: Comparison of UV C light and chemicals for disinfection of surfaces in hospital isolation rooms. Infect Control Hosp Epidemiol 2006, 27:729-734.
- [9]Nerandzic MM, Cadnum JL, Pultz MJ, Donskey CJ: Evaluation of an automated ultraviolet radiation device for decontamination of Clostridium difficile and other healthcare-associated pathogens in hospital rooms. BMC Infect Dis 2010, 10:197-204. BioMed Central Full Text
- [10]Rutala WA, Gergen MF, Weber DJ: Room decontamination with UV radiation. Infect Control Hosp Epidemiol 2010, 31:1025-1029.
- [11]Rastogi VK, Wallace L, Smith LS: Disinfection of Acinetobacter baumannii-contaminated surfaces relevant to medical treatment facilities with ultraviolet C light. Military Med 2007, 11:1166-1169.
- [12]Shetty N, Srinivasan S, Holton J, Ridgway GL: Evaluation of microbicidal activity of a new disinfectant: Sterilox 2500 against Clostridium difficile spores, Helicobacter pylori, vancomycin resistant Enterococcus species, Candida albicans and several Mycobacterium species. J Hosp Infect 1999, 41:101-105.
- [13]Heim A, Ebnet C, Harste G, Pring-Akerblom P: Rapid and quantitative detection of human Adenovirus DNA by real-time PCR. J Med Virol 2003, 70:228-239.
- [14]Mitchell A, DeAlwis N, Collins J, Chew K, Taylor R, Schwa U, Narayanan M: Stethoscope or ‘staphoscope’? Infection by auscultation. J Hosp Infect 2010, 76:278-279.
- [15]Porwancher R, Sheth A, Remphrey S, Taylor E, Hinkle C, Zervos M: Epidemiological study of hospital-acquired infection with vancomycin-resistant Enterococcus faecium: Possible transmission by an electronic ear-probe thermometer. Infect Control Hosp Epidemiol 1997, 18:771-773.
- [16]Rutala WA, White MS, Gergen MF, Weber DJ: Bacterial contamination of keyboards: efficacy and functional impact of disinfectants. Infect Control Hosp Epidemiol 2006, 27:372-377.
- [17]Cefai C, Elliot TSJ: Re-usable ECG electrodes – a vehicle for cross-infection? J Hosp Infect 1988, 12:65-67.
- [18]Creamer E, Humphreys H: The contribution of beds to healthcare-associated infection: the importance of adequate decontamination. J Hosp Infect 2008, 69:8-23.
- [19]Sattar SA: Promises and pitfalls of recent advances in chemical means of preventing the spread of nosocomial infections by environmental surfaces. Am J Infect Control 2010, 38:S34-S40.
- [20]Casey AL, Adams D, Karpanen TJ, Lambert PA, Cookson BD, Nightingale P, Miruszenko L, Shillam R, Christian P, Elliott TSJ: Role of copper in reducing hospital environment contamination. J Hosp Infect 2010, 74:72-77.
- [21]Otter JA, Cummins M, Ahmad F, van Tonder C, Drabu YJ: Assessing the biological efficacy and rate of recontamination following hydrogen peroxide vapour decontamination. J Hosp Infect 1997, 67:182-188.
- [22]Medicine and Healthcare Products Regulatory Agency: Medical device alert (MDA/2007/034). , London; 2007.
- [23]Gritz DC, Lee TY, McDonnell PJ, Shih K, Baron N: Ultraviolet radiation for the sterilization of contact lenses. Eye Contact Lens 1990, 16:294-298.
- [24]Büttgen S, Gebel J, Hornei B, Engelhart S, Koch ONJ, Exner M: Comparison of chemo resistance of Clostridium difficile ribotype 027 spores and Bacillus subtilis spores against disinfectants. Hyg Med 2008, 33:513-517.
- [25]Snelling AM, Beggs CB, Kerr KG, Shepherd SJ: Spores of Clostridium difficile in hospital air. Clin Infect Dis 2010, 51:1104-1105.
- [26]Moisan M, Barbeau J, Moreau S, Pelletier J, Tabrizian M, Yahia LH: Low-temperature sterilization using gas plasmas: a review of the experiments and an analysis of the inactivation mechanisms. Int J Pharm 2001, 226:1-21.
- [27]Humphreys PN: Testing standards for sporicides. J Hosp Infect 2011, 77:193-198.
- [28]Hijnen WAM, Beerendonk EF, Medema GJ: Inactivation credit of UV radiation for viruses, bacteria and protozoan (oo)cysts in water. Water Res 2006, 40:3-22.
PDF