Fluoroscopic views for safe insertion of lag screws into the posterior column of the acetabulum
Yong Shen1  Yingze Zhang1  Jia Li1  Yang Lu1  Zekun Zhang2  Wei Chen1 
[1] Department of Orthopaedic Surgery, Third Hospital of Hebei Medical University, Shijiazhuang, Hebei 050051, P.R. China;Department of Radiology, Third Hospital of Hebei Medical University, Shijiazhuang, Hebei 050051, P.R. China
关键词: Fluoroscopic view;    Percutaneous fixation;    Posterior column;    Acetabular fracture;   
Others  :  1122116
DOI  :  10.1186/1471-2474-15-303
 received in 2014-03-15, accepted in 2014-08-28,  发布年份 2014
PDF
【 摘 要 】

Background

Percutaneous lag screw fixation is an alternative treatment for non-displaced or minimally displaced posterior column fractures. This study aims to explore new fluoroscopic views of the acetabulum for safe percutaneous insertion of posterior column lag screws.

Methods

Axial computed tomography (CT) scans were taken of sixteen embalmed adult cadavers. The axial CT images at the level of the middle height of the acetabulum were selected. The angle (angle α) between the posterior cortex of the posterior column (PCPC) and the line intersecting the axial plane and the coronal plane, and the angle (angle β) between the medial wall and the line intersecting the axial plane and the sagittal plane were identified and measured. Tangential views of the PCPC and medial wall were obtained by referencing the measured angles. A lag screw was inserted into the posterior columns of the sixteen pelvic specimens under fluoroscopic guidance using an iliac oblique view and the two tangential views. CT scans were performed to evaluate the lag screw position. Axial CT images of 52 volunteers were obtained and the angles α and β were measured following the same methods used for the cadaveric specimens.

Results

The angles α and β for the specimens were 29.3±2.8.1 and 8.1±1.4 degrees, respectively. On the tangential view of the PCPC, the posterior cortex appears as a nearly straight line between the lesser and greater sciatic notches. On the tangential view of the medial wall, the medial wall appears as a distinct straight line. Using these radiographic images, the lag screws were inserted into the posterior columns of bony pelvic specimens. Screw placement was confirmed by CT, and found to be fully intraosseous in all cases without any cortical breaches. The angles α and β were 30.4±4.1 and 9.2±1.9 degrees for male volunteers and 28.5±3.7 and 7.7±1.8 degrees for female volunteers, significant difference in these angles between cadaveric specimens and human volunteers.

Conclusion

The tangential views of both the PCPC and medial wall can be obtained following the aforementioned methods The oblique iliac view and the two tangential views enable safe insertion of posterior column lag screws.

【 授权许可】

   
2014 Chen et al.; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20150213022810268.pdf 1218KB PDF download
Figure 6. 56KB Image download
Figure 5. 84KB Image download
Figure 4. 79KB Image download
Figure 3. 69KB Image download
Figure 2. 118KB Image download
Figure 1. 117KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

【 参考文献 】
  • [1]Mouhsine E, Garofalo R, Borens O, Wettstein M, Blanc CH, Fischer JF, Moretti B, Leyvraz PF: Percutaneous retrograde screwing for stabilisation of acetabular fractures. Injury 2005, 36(11):1330-1336.
  • [2]Parker PJ, Copeland C: Percutaneous fluoroscopic screw fixation of acetabular fractures. Injury 1997, 28(9–10):597-600.
  • [3]Sanders MB, Starr AJ, Reinert C: Percutaneous screw fixation of acetabular fractures in elderly patients. Curr Opin Orthop 2006, 17(1):17-24.
  • [4]Starr AJ, Reinert CM, Jones AL: Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma 1998, 12(1):51-58.
  • [5]Gay SB, Sistrom C, Wang GJ, Kahler DA, Boman T, McHugh N, Goitz HT: Percutaneous screw fixation of acetabular fractures with CT guidance: preliminary results of a new technique. AJR Am J Roentgenol 1992, 158(4):819-822.
  • [6]Starr AJ, Jones AL, Reinert CM, Borer DS: Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury 2001, 32(Suppl 1):SA45-SA50.
  • [7]Zura RD, Kahler DM: A transverse acetabular nonunion treated with computer-assisted percutaneous internal fixation. A case report. J Bone Joint Surg Am 2000, 82(2):219-224.
  • [8]Brown GA, Willis MC, Firoozbakhsh K, Barmada A, Tessman CL, Montgomery A: Computed tomography image-guided surgery in complex acetabular fractures. Clin Orthop Relat Res 2000, 370:219-226.
  • [9]Mears DC: Surgical treatment of acetabular fractures in elderly patients with osteoporotic bone. J Am Acad Orthop Surg 1999, 7(2):128-141.
  • [10]Templeman DC, Olson S, Moed BR, Duwelius P, Matta JM: Surgical treatment of acetabular fractures. Instr Course Lect 1999, 48:481-496.
  • [11]Shahulhameed A, Roberts CS, Pomeroy CL, Acland RD, Giannoudis PV: Mapping the columns of the acetabulum–implications for percutaneous fixation. Injury 2010, 41(4):339-342.
  • [12]Yang JH, Chouhan DK, Oh KJ: Percutaneous screw fixation of acetabular fractures: applicability of hip arthroscopy. Arthroscopy 2010, 26(11):1556-1561.
  • [13]Mosheiff R, Khoury A, Weil Y, Liebergall M: First generation computerized fluoroscopic navigation in percutaneous pelvic surgery. J Orthop Trauma 2004, 18(2):106-111.
  • [14]Geijer M, El-Khoury GY: Imaging of the acetabulum in the era of multidetector computed tomography. Emerg Radiol 2007, 14(5):271-287.
  • [15]Carmack DB, Moed BR, McCarroll K, Freccero D: Accuracy of detecting screw penetration of the acetabulum with intraoperative fluoroscopy and computed tomography. J Bone Joint Surg Am 2001, 83-A(9):1370-1375.
  • [16]Letournel E, Judet R: Radiology of the Normal Acetabulum. In Fractures of the Acetabulum. 2nd edition. Edited by Letournel E, Judet R, Elson RA. Berlin: Springer-Verlag; 1993:29-61.
  • [17]Magid D, Fishman EK, Brooker AF Jr, Mandelbaum BR, Siegelman SS: Multiplanar computed tomography of acetabular fractures. J Comput Assist Tomogr 1986, 10(5):778-783.
  • [18]Mandelbaum BR, Magid D, Fishman EK, Burgess AR, Bosse M, Brooker AF Jr: Multiplanar computed tomography: a multidimensional tool for evaluation and treatment of acetabular fractures. J Comput Tomogr 1987, 11(2):167-173.
  • [19]Matta JM: Fractures of the acetabulum: accuracy of reduction and clinical results in patients managed operatively within three weeks after the injury. J Bone Joint Surg Am 1996, 78(11):1632-1645.
  • [20]Olson SA, Matta JM: The computerized tomography subchondral arc: a new method of assessing acetabular articular continuity after fracture (a preliminary report). J Orthop Trauma 1993, 7(5):402-413.
  • [21]White MS: Three-dimensional computed tomography in the assessment of fractures of the acetabulum. Injury 1991, 22(1):13-19.
  • [22]Wright R, Barrett K, Christie MJ, Johnson KD: Acetabular fractures: long-term follow-up of open reduction and internal fixation. J Orthop Trauma 1994, 8(5):397-403.
  • [23]Wang Q, Zhang Y, Pan J, Peng A, Wu X, Wang P, Li H, Zhang S, Zhang X, Zhao L: Applied anatomy of retrograde lag screw internal fixation in the posterior column of the acetabulum. Chin J Clin Anat 2006, 24(1):36-38.
  • [24]Kellam JF, Tile M: Surgical Techniques. In Fractures of the Pelvis and Acetabulum. 2nd edition. Edited by Tile M. Baltimore: Williams and Wilkins; 1995:355-396.
  • [25]Ebraheim NA, Savolaine ER, Hoeflinger MJ, Jackson WT: Radiological diagnosis of screw penetration of the hip joint in acetabular fracture reconstruction. J Orthop Trauma 1989, 3(3):196-201.
  • [26]Mears DC, MacLeod MD: Acetabular Fractures: Triradiate and Modified Triradiate Approaches. In Master Techniques in Orthopaedic Surgery: Fractures. Edited by Wiss DA. Philadelphia: Lippincott-Raven; 1998:697-724.
  • [27]Helfet DL, Bartlett CS, Malkani AL: Acetabular Fractures: Extended Iliofemoral Approach. In Master Techniques in Orthopaedic Surgery: Fractures. Edited by Wiss DA. Philadelphia: Lippincott-Raven; 1998:675-696.
  • [28]Norris BL, Hahn DH, Bosse MJ, Kellam JF, Sims SH: Intraoperative fluoroscopy to evaluate fracture reduction and hardware placement during acetabular surgery. J Orthop Trauma 1999, 13(6):414-417.
  • [29]Kahler DM: Navigated long-bone fracture reduction. J Bone Joint Surg Am 2009, 91(Suppl 1):102-107.
  • [30]Gras F, Marintschev I, Klos K, Muckley T, Hofmann GO, Kahler DM: Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study. J Orthop Trauma 2012, 26(8):466-473.
  • [31]Gras F, Marintschev I, Mendler F, Wilharm A, Muckley T, Hofmann GO: 2D-fluoroscopic navigated screw osteosynthesis of acetabular fractures: a preliminary report. Z Orthop Unfall 2008, 146(2):231-239.
  文献评价指标  
  下载次数:58次 浏览次数:13次