Novel biaxial tensile test for studying aortic failure phenomena at a microscopic level
Shukei Sugita1  Takeo Matsumoto2 
[1] Center for Fostering Young and Innovative Researchers, Nagoya Institute of Technology, Gokiso-cho, Showa-ku, Nagoya, 466-8555, Japan
[2] Department of Mechanical Engineering, Graduate School of Engineering, Nagoya Institute of Technology, Gokiso-cho, Showa-ku, Nagoya, 466-8555, Japan
关键词: Tearing;    Microstructure;    Biaxial tensile device;    Aortic tissue;   
Others  :  797961
DOI  :  10.1186/1475-925X-12-3
 received in 2012-09-10, accepted in 2012-12-26,  发布年份 2013
PDF
【 摘 要 】

Background

An aortic aneurysm is a local dilation of the aorta, which tends to expand and often results in a fatal rupture. Although larger aneurysms have a greater risk of rupture, some small aneurysms also rupture. Since the mechanism of aortic rupture is not well understood, clarification of the microstructure influencing the failure to rupture is important. Since aortic tissues are stretched biaxially in vivo, we developed a technique to microscopically observe the failure of an aortic rupture during biaxial stretch.

Methods

A thinly sliced porcine thoracic aortic specimen was adhered to a circular frame and pushed onto a cylinder with a smaller diameter to stretch the specimen biaxially. To induce failure to rupture at the center, the specimen was thinned at the center of the hole as follows: the specimen was frozen while being compressed with metal plates having holes, which were 3 mm in diameter at their centers; the specimen was then sliced at 50-μm intervals and thawed.

Results

The ratio of the thickness at the center to the peripheral area was 99.5% for uncompressed specimens. The ratio decreased with an increase in the compression ratio εc and was 47.3% for specimens with εc = 40%. All specimens could be stretched until failure to rupture. The probability for crack initiation within the cylinder was <30% and 100% for specimens with εc <10% and εc >30%, respectively. Among specimens ruptured within the cylinder, 93% of those obtained from the mid-media showed crack initiation at the thin center area.

Conclusions

Aortic tissues were successfully stretched biaxially until failure, and their crack initiation points were successfully observed under a microscope. This could be a very useful and powerful method for clarifying the mechanism of aortic rupture. We are planning to use this technique for a detailed investigation of events occurring at the point of failure when the crack initiates in the aortic aneurysm wall.

【 授权许可】

   
2013 Sugita and Matsumoto; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140706092032553.pdf 1505KB PDF download
Figure 8. 58KB Image download
Figure 7. 78KB Image download
Figure 6. 77KB Image download
Figure 5. 47KB Image download
Figure 4. 122KB Image download
Figure 3. 120KB Image download
Figure 2. 146KB Image download
Figure 1. 59KB Image download
【 图 表 】

Figure 1.

Figure 2.

Figure 3.

Figure 4.

Figure 5.

Figure 6.

Figure 7.

Figure 8.

【 参考文献 】
  • [1]Johansson G, Markstrom U, Swedenborg J: Ruptured thoracic aortic aneurysms: a study of incidence and mortality rates. J Vasc Surg 1995, 21:985-988.
  • [2]Coady MA, Rizzo JA, Hammond GL, Mandapati D, Darr U, Kopf GS, Elefteriades JA: What is the appropriate size criterion for resection of thoracic aortic aneurysms? J Thorac Cardiovasc Surg 1997, 113:476-491.
  • [3]Cambria RA, Gloviczki P, Stanson AW, Cherry KJ Jr, Bower TC, Hallett JW Jr, Pairolero PC: Outcome and expansion rate of 57 thoracoabdominal aortic aneurysms managed nonoperatively. Am J Surg 1995, 170:213-217.
  • [4]Sokolis DP, Kritharis EP, Giagini AT, Lampropoulos KM, Papadodima SA, Iliopoulos DC: Biomechanical response of ascending thoracic aortic aneurysms: association with structural remodelling. Comput Meth Biomech Biomed Eng 2012, 15:231-248.
  • [5]Iliopoulos DC, Kritharis EP, Giagini AT, Papadodima SA, Sokolis DP: Ascending thoracic aortic aneurysms are associated with compositional remodeling and vessel stiffening but not weakening in age-matched subjects. J Thorac Cardiovasc Surg 2009, 137:101-109.
  • [6]Garcia-Herrera CM, Atienza JM, Rojo FJ, Claes E, Guinea GV, Celentano DJ, Garcia-Montero C, Burgos RL: Mechanical behaviour and rupture of normal and pathological human ascending aortic wall. Med Biol Eng Comput 50:559-566.
  • [7]Isselbacher EM: Thoracic and abdominal aortic aneurysms. Circulation 2005, 111:816-828.
  • [8]He CM, Roach MR: The composition and mechanical properties of abdominal aortic aneurysms. J Vasc Surg 1994, 20:6-13.
  • [9]Fukui T, Matsumoto T, Tanaka T, Ohashi T, Kumagai K, Akimoto H, Tabayashi K, Sato M: In vivo mechanical properties of thoracic aortic aneurysmal wall estimated from in vitro biaxial tensile test. Biomed Mater Eng 2005, 15:295-305.
  • [10]Matsumoto T, Fukui T, Tanaka T, Ikuta N, Ohashi T, Kumagai K, Akimoto H, Tabayashi K, Sato M: Biaxial tensile properties of thoracic aortic aneurysm tissues. J Biomech Sci Eng 2009, 4:518-529.
  • [11]Okamoto RJ, Wagenseil JE, DeLong WR, Peterson SJ, Kouchoukos NT, Sundt TM: 3rd: Mechanical properties of dilated human ascending aorta. Ann Biomed Eng 2002, 30:624-635.
  • [12]Marra SP, Kennedy FE, Kinkaid JN, Fillinger MF: Elastic and rupture properties of porcine aortic tissue measured using inflation testing. Cardiovasc Eng 2006, 6:123-131.
  • [13]Mohan D, Melvin JW: Failure properties of passive human aortic tissue. II--Biaxial tension tests. J Biomech 1983, 16:31-44.
  • [14]Ohashi T, Sugita S, Matsumoto T, Kumagai K, Akimoto H, Tabayashi K, Sato M: Rupture properties of blood vessel walls measured by pressure-imposed test. JSME Int J Ser C 2003, 46:1290-1296.
  • [15]Sugita S, Matsumoto T, Ohashi T, Kumagai K, Akimoto H, Tabayashi K, Sato M: Evaluation of rupture properties of thoracic aortic aneurysms in a pressure-imposed test for rupture risk estimation. Cardiovasc Eng Technol 2011, 3:41-51.
  • [16]Wicker BK, Hutchens HP, Wu Q, Yeh AT, Humphrey JD: Normal basilar artery structure and biaxial mechanical behaviour. Comput Meth Biomech Biomed Eng 2008, 11:539-551.
  • [17]Chen H, Liu Y, Slipchenko MN, Zhao X, Cheng JX, Kassab GS: The layered structure of coronary adventitia under mechanical load. Biophys J 2011, 101:2555-2562.
  • [18]Huang L, Mathieu PS, Helmke BP: A stretching device for high-resolution live-cell imaging. Ann Biomed Eng 2010, 38:1728-1740.
  • [19]Hung CT, Williams JL: A method for inducing equi-biaxial and uniform strains in elastomeric membranes used as cell substrates. J Biomech 1994, 27:227-232.
  • [20]Meijering E, Dzyubachyk O, Smal I: Methods for cell and particle tracking. Methods Enzymol 2012, 504:183-200.
  • [21]Otsu N: A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cybern 1979, 9:62-66.
  • [22]Sugita S, Matsumoto T: Quantitative measurement of the distribution and alignment of collagen fibers in unfixed aortic tissues. J Biomech 2012. submitted
  • [23]Langerak SE, Groenink M, van der Wall EE, Wassenaar C, Vanbavel E, van Baal MC, Spaan JA: Impact of current cryopreservation procedures on mechanical and functional properties of human aortic homografts. Transpl Int 2001, 14:248-255.
  • [24]Goh KL, Chen Y, Chou SM, Listrat A, Bechet D, Wess TJ: Effects of frozen storage temperature on the elasticity of tendons from a small murine model. Animal 2010, 4:1613-1617.
  • [25]Vaishnav RN, Vossoughi J: Residual stress and strain in aortic segments. J Biomech 1987, 20:235-239.
  • [26]Guo X, Kassab GS: Distribution of stress and strain along the porcine aorta and coronary arterial tree. Am J Physiol Heart Circ Physiol 2004, 286:H2361-H2368.
  • [27]Matsumoto T, Hayashi K, Ide K: Residual strain and local strain distributions in the rabbit atherosclerotic aorta. J Biomech 1995, 28:1207-1217.
  文献评价指标  
  下载次数:54次 浏览次数:25次