Surface pretreatment for prolonged survival of cemented tibial prosthesis components: full- vs. surface-cementation technique
Thorsten Mumme1  Fritz Uwe Niethard1  Dieter Christian Wirtz1  Mutaz Qunaibi2  Rudolf Marx2 
[1]Department of Orthopedic Surgery, University Hospital of the University of Technology, Aachen, Germany
[2]Department of Prosthetic Dentistry, Section of Dental Materials, University Hospital of the University of Technology, Aachen, Germany
Others  :  798786
DOI  :  10.1186/1475-925X-4-61
 received in 2005-05-09, accepted in 2005-10-31,  发布年份 2005
PDF
【 摘 要 】

Background

One of few persisting problems of cemented total knee arthroplasty (TKA) is aseptic loosening of tibial component due to degradation of the interface between bone cement and metallic tibial shaft component, particularly for surface cemented tibial components. Surface cementation technique has important clinical meaning in case of revision and for avoidance of stress shielding. Degradation of the interface between bone cement and bone may be a secondary effect due to excessive crack formation in bone cement starting at the opposite metallic surface.

Methods

This study was done to prove crack formation in the bone cement near the metallic surface when this is not coated. We propose a newly developed coating process by PVD layering with SiOx to avoid that crack formation in the bone cement. A biomechanical model for vibration fatigue test was done to simulate the physiological and biomechanical conditions of the human knee joint and to prove excessive crack formation.

Results

It was found that coated tibial components showed a highly significant reduction of cement cracking near the interface metal/bone cement (p < 0.01) and a significant reduction of gap formation in the interface metal-to-bone cement (p < 0.05).

Conclusion

Coating dramatically reduces hydrolytic- and stress-related crack formation at the prosthesis interface metal/bone cement. This leads to a more homogenous load transfer into the cement mantle which should reduce the frequency of loosening in the interfaces metal/bone cement/bone. With surface coating of the tibial component it should become possible that surface cemented TKAs reveal similar loosening rates as TKAs both surface and stem cemented. This would be an important clinical advantage since it is believed that surface cementing reduces metaphyseal bone loss in case of revision and stress shielding for better bone health.

【 授权许可】

   
2005 Marx et al; licensee BioMed Central Ltd.

【 预 览 】
附件列表
Files Size Format View
20140706183644998.pdf 1087KB PDF download
Figure 6. 33KB Image download
Figure 5. 22KB Image download
Figure 4. 112KB Image download
Figure 1. 38KB Image download
Figure 2. 46KB Image download
Figure 3. 155KB Image download
【 图 表 】

Figure 3.

Figure 2.

Figure 1.

Figure 4.

Figure 5.

Figure 6.

【 参考文献 】
  • [1]Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM: Why Are Total Knee Arthroplasties Failing Today? Clin Orthop 2002, 404:7-13.
  • [2]Sharkey PF, Hozack WJ, Rothman RH, Shastri S, Jacoby SM: Why Are Total Knee Arthroplasties Failing Today? John Insall Award. Program and abstracts of the Combined Speciality Day meeting of the AAHKS. Dallas, Texas; February 16, 2002
  • [3]Austin MS, Sharkey PF, Hozack WJ, Rothman RH: Knee Failure Mechanisms After Total Knee Arthroplasty. Techniques in Knee Surgery 2004, 3:55-59.
  • [4]Gonzalez MH, Mekhail AO: The Failed Total Knee Arthroplasty: Evaluation and Etiology. J Am Acad Orthop Surg 2004, 12:436-446.
  • [5]Rand JA, Trousdale RT, Ilstrup DM, Harmsen WS: Factors Affecting the Durability of Primary Total Knee Prostheses. J Bone Joint Surg Am 2003, 85:259-265.
  • [6]Scuderi GR, Pagnano MW: The rationale for posterior cruciate substituting total knee arthroplasty. J Orthop Surg 2001, 9:81-88.
  • [7]Gill GS, Joshi AB: Long-term results of Kinematic Condylar knee replacement. J Bone Joint Surg [Br] 2001, 83-B:355-358.
  • [8]Peters CL, Craig MA, Mohr RA, Bachus KN: Tibial Component Fixation With Cement. Full-Versus Surface-Cementation Techniques. Clin Orthop 2003, 409:158-168.
  • [9]Van Loon CJ, de Waal Malefijt MC, Buma P, Verdonschot N, Veth RP: Femoral bone loss in total knee arthroplasty. A review. Acta Orthop Belg 1999, 65:154-163.
  • [10]Nyman JS, Hazelwood SJ, Rodrigo JJ, Martin RB, Yeh OC: Long stemmed total knee arthroplasty with interlocking screws: a computational bone adaption study. J Orthop Res 2004, 22:51-57.
  • [11]Spiegelberg StH, Ruberti JW, Braithwaite GGC: System and methods for reducing interfacial porosity in cements. US Patent 6,884,264 issued on 04/26/2005
  • [12]Green DJ: An introduction to the mechanical properties of ceramics. Cambridge, Cambridge University Press; 1998.
  • [13]Munz D, Fett T: Ceramics – Mechanical Properties, Failure Behaviour, Materials Selection. Berlin Heidelberg, Springer; 1999.
  • [14]Erli HJ, Marx R, Paar O, Niethard FU, Weber M, Wirtz DC: Surface pretreatments for medical application of adhesion. BioMedical Engineering OnLine 2003, 2:15-33. BioMed Central Full Text
  • [15]Plueddemann EP: Silane coupling agents. New York, Plenum Press; 1982.
  • [16]Kinloch AJ: Adhesion and Adhesives. London, Chapman and Hall; 1987.
  • [17]Habenicht G: Kleben. Grundlagen, Technologien, Anwendungen. Berlin, Springer; 1990.
  • [18]Brewis DM, Comyn J, Tegg JL: The durability of some epoxide adhesive-bonded joints on exposure to moist warm air. Int J Adhesion and Adhesives 1980, 1:35-39.
  • [19]Marx R: Ursachen für das Versagen von Klebeverbindungen bei Adhäsivbrücken. Schweiz Monatssch Zahnmed 1987, 97:1081-1086.
  • [20]Marx R: Auswahl der Werkstoffe. In Adhäsivprothetik München, Urban und Schwarzenberg Edited by Kerschbaum Th. 1995, 107-120.
  • [21]Ritter MA, Herbst SA, Keating EM, Faris PM: Radiolucency at the bone-cement interface in total knee replacement. J Bone Joint Surg 1994, 76A:60-65.
  • [22]Ecker ML, Lotke PA, Windsor RE, Cella JP: Long-term results after total condylar knee arthroplasty. Significance of radiolucent lines. Clin Orthop 1987, 216:151-158.
  • [23]Musil R, Tiller HJ: Der Kunststoff-Metall-Verbund in der zahnärztlichen Prothetik. Leipzig, VEB J.A. Barth 1988.
  • [24]Marx R, Fischer H: Work piece and method for producing and utilizing said work piece. US Patent 6,777,028 issued on 08/17/2004
  • [25]Wirtz DC: Eine neue Beschichtungsmethode für zementierte Femurschaftimplantate zur hydrolysestabilen Optimierung des Metall-Knochenzement-Verbundes. Aachen, Aachener Beiträge zur Medizin, Band 25, Wissenschaftsverlag Mainz, Aachen 2002. ISBN 3-86073-945-x
  • [26]Bergmann G: In vivo Messung der Belastung von Hüftimplantaten. Habilitationsschrift, Freie Universität Berlin 1994.
  • [27]Ong A, Wong KL, Lai M, Garino JP, Steinberg ME: Early failure of precoated femoral components in primary total hip arthroplasty. J Bone Joint Surg Am 2002, 84:786-792.
  • [28]Strickland AB, Chen YH, Andriacchi TP, Miller J: The initial fixation of porous coated tibial components evaluated by the study of rigid body motion under static load. Trans Orthop Res Soc 1988, 13:476-481.
  文献评价指标  
  下载次数:91次 浏览次数:97次