| 2018 2nd International Conference on Advanced Technologies in Manufacturing and Materials Engineering | |
| Experimental Study and Matrix Effect Correction of Pseudobinary Samples in XRF Analysis | |
| Liu, Hefan^1 ; Zhou, Xiaoling^1 ; Zhang, Xin^1 ; Wu, Keyin^1 ; Lu, Chengwei^1 | |
| Chengdu Research Academy of Environmental Sciences, Chengdu, China^1 | |
| 关键词: Calibration procedure; Content measurements; Correction method; Influence coefficient; Maximum absolute error; Relative errors; Si pin-detector; X ray fluorescence analysis; | |
| Others : https://iopscience.iop.org/article/10.1088/1757-899X/389/1/012002/pdf DOI : 10.1088/1757-899X/389/1/012002 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
This basic study conducted an X-ray fluorescence (XRF) analysis on pseudo-binary samples, which emphasis on the binary influence coefficient and correction for matrix effects. The Ti-V, V-Ti, Ti-Fe, Fe-Ti, V-Fe and Fe-V samples were confected and measured by XRF analysis using a Si-PIN detector. These coefficients were calculated using mathematical models proposed through the Lachance-Traill algorithm, which can be applied in practice via an efficient calibration procedure. For the content measurement, the maximum absolute error decreased from 5.92% to 0.27%, and the relative error decreased from 14.84% to 0.66% via fitting correction for matrix effects. This study describes and emphasizes the application of the binary influence coefficient and fitting correction method.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Experimental Study and Matrix Effect Correction of Pseudobinary Samples in XRF Analysis | 711KB |
PDF