| International Conference on Applied Sciences 2017 | |
| Maintenance of reference standards in the field of viscosity | |
| Moulic, E.A.^1 ; Cirneanu, I.^1 ; Constantin, N.^2 ; Rucai, V.^2 | |
| National Institute of Metrology (INM), Department of Physico-chemical Quantities, Bucharest, Romania^1 | |
| University Politehnica of Bucharest, Department of Material Science, Bucharest, Romania^2 | |
| 关键词: American society for testing and materials; Flow properties; Newtonian fluids; Operation programs; Reference standard; | |
| Others : https://iopscience.iop.org/article/10.1088/1757-899X/294/1/012019/pdf DOI : 10.1088/1757-899X/294/1/012019 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
Participation in the work of comparison in the field of viscosity, within the program conducted under the jurisdiction of ASTM (American Society for Testing and Materials), D-2 Committee, Subcommittee Flow Properties, Newtonian Fluids) was necessary to ensure traceability of measuring unit of kinematic viscosity. Results of the comparison of the specialized participating laboratories on 4 continents, has proved annual capability of INM in the transmission unit of kinematic viscosity. Cannon Position Company in the US organizes co-operation program in the field of kinematic viscosity ASTM D 02.07. The company distributes standard substances Cannon viscosity participating laboratories and consolidate the results of the measurements. Physical-chemical laboratory has fully accepted the proposed schedule of the company Cannon. Final report of the comparison showed that in the year 2015 a number of 25 laboratories and institutes of metrology attented to the program.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Maintenance of reference standards in the field of viscosity | 577KB |
PDF