会议论文详细信息
10th International School-Conference on Materials for Extreme Environment: Development, Production and Application
Two-ply anode X-ray tube for computed tomography scanner
Ignatyev, D.^1,2 ; Taubin, M.^1,2 ; Chesnokov, D.^1 ; Malyshev, V.^1 ; Yaskolko, A.^1,2
Federal State Unitary Enterprise, Science Research Institute, LUCH, 24 Zheleznodorozhnaya st, Podolsk, Moscow region
142100, Russia^1
National Research Nuclear University MEPhI, Moscow Engineering Physics Institute, Kashirskoe highway, 31, Moscow
115409, Russia^2
关键词: Computed tomography scanners;    Computer tomographs;    Graphite surfaces;    Thermal field;    Tungsten coating;    Tungsten layers;   
Others  :  https://iopscience.iop.org/article/10.1088/1757-899X/130/1/012035/pdf
DOI  :  10.1088/1757-899X/130/1/012035
来源: IOP
PDF
【 摘 要 】

This report presents a method of the formation of tungsten layer on the graphite surface. The described method can be used to create the anode of powerful x-ray tubes for medical purposes, in particular, a computer tomograph (CT). The thermal properties of the graphite base and the deposited tungsten coating, as well as the strength of the resulting coating were studied. Thermal fields in the CT-anode with a power of 100 kW were calculated.

【 预 览 】
附件列表
Files Size Format View
Two-ply anode X-ray tube for computed tomography scanner 1474KB PDF download
  文献评价指标  