International Conference on Polarised Neutrons for Condensed Matter Investigations | |
Optimization of a polarizer device for SANS-2 instrument at the PIK reactor | |
Pavlov, K.A.^1,2 ; Konik, P.I.^1,2 ; Syromyatnikov, V.G.^1,2 ; Grigoriev, S.V.^1,2 ; Moskvin, E.V.^1,2 | |
Petersburg Nuclear Physics Institute NRC KI, Gatchina, 188300, Russia^1 | |
Saint Petersburg State University, St-Petersburg | |
199034, Russia^2 | |
关键词: Beam polarization; Device performance; Input parameter; Supermirrors; Wavelength ranges; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/862/1/012020/pdf DOI : 10.1088/1742-6596/862/1/012020 |
|
来源: IOP | |
【 摘 要 】
A polarizer of V-cavity type for small angle diffractometer was considered. The device performance was simulated using McStas ray tracing package with the input parameters for supermirrors of PNPI home production (CoFe/TiZr, m = 2.13). Two replaceable devices with lengths 0.75 m and 1.8 m are proposed to cover the wavelength range 4-25 Å with beam polarization not less than 95%.