会议论文详细信息
International Conference on Polarised Neutrons for Condensed Matter Investigations
Thermal stabilization of neutron Larmor diffractometers
Keller, T.^1,2 ; Tralmer, F.^1,2
Max Planck Institute for Solid State Research, Stuttgart
D-70569, Germany^1
Max Planck Society Outstation at the FRM II, Garching
D-85748, Germany^2
关键词: Accuracy level;    Low thermal expansion;    Radio frequency coils;    RF coil;    Support structures;    Thermal stabilization;    Zerodur;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/862/1/012012/pdf
DOI  :  10.1088/1742-6596/862/1/012012
来源: IOP
PDF
【 摘 要 】

We report on the design of a support unit for the radio frequency (RF) coils of a Larmor diffractometer (LD) eliminating fluctuations of the Larmor phase resulting from thermal expansion of the support structures. The key component defining the spacing between the RF coils is a Zerodur bar with a very low thermal expansion coefficient (α = 7 × 10-8K-1). This support unit will allow for LD measurements on the 10-6accuracy level even if the ambient temperature is fluctuating.

【 预 览 】
附件列表
Files Size Format View
Thermal stabilization of neutron Larmor diffractometers 2932KB PDF download