| X-Ray Microscopy Conference 2016 | |
| Introducing high efficiency image detector to X-ray imaging tomography | |
| Uesugi, Kentaro^1 ; Hoshino, Masato^1 ; Takeuchi, Akihisa^1 | |
| Japan Synchrotron Radiation Research Institute, JASRI/SPring-8, Kouto 1-1-1, Sayo, Hyogo | |
| 679-5198, Japan^1 | |
| 关键词: High reproducibility; High-efficiency; Linear absorption coefficient; Micro-tomography; Phosphor screens; Straight fibers; X ray intensity; X-ray image detectors; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/849/1/012051/pdf DOI : 10.1088/1742-6596/849/1/012051 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
A high efficiency X-ray image detector was introduced in an x-ray imaging micro-tomography system. The detector is a visible light conversion type which consists of phosphor screen, straight-fiber (1:1) optics and scientific CMOS device. The obtained CT images show high reproducibility of X-ray linear absorption coefficient of test material (Cu/Al concentric pattern). The x-ray intensity could be reduced about 10 % of previous system with lens-coupled image detector system.