会议论文详细信息
| 9th International Conference on 3D Radiation Dosimetry | |
| Clinical management of tumour volume changes in VMAT head & neck radiation treatment | |
| Alexander, K.M.^1 ; Gooding, J.^2 ; Schreiner, L.J.^1,2 ; Olding, T.^1,2 | |
| Department of Physics, Queen's University, Kingston | |
| ON | |
| K7L 3N6, Canada^1 | |
| Cancer Centre of South Eastern Ontario, Kingston General Hospital, 25 King Street West, Kingston | |
| ON | |
| K7L 5P9, Canada^2 | |
| 关键词: Body contours; Clinical management; Gel dosimeter; Radiation treatments; Variable sizes; Verification plans; Volume change; Xylenol Orange; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/847/1/012038/pdf DOI : 10.1088/1742-6596/847/1/012038 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
The impact of changing anatomy due to tumour shrinkage was assessed for a VMAT head & neck plan optimized according to our centre's planning protocol. A custom-built wax phantom accommodating ion chamber, MOSFET, EBT3 film and Fricke-xylenol orange-gelatin (FXG) gel dosimeters and a variable size bolus 'tumour' was used in the investigation. Results indicate that the practice of initiating a patient re-scan and verification plan is appropriate when a change in external body contour greater than 1 cm compared to the original anatomy is observed.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Clinical management of tumour volume changes in VMAT head & neck radiation treatment | 340KB |
PDF