会议论文详细信息
9th International Conference on 3D Radiation Dosimetry
Quantifying refractive index mismatch effects on cone beam optical CT scanner measurements
Carpentier, E.E.^1 ; Alexander, K.M.^1 ; Schreiner, L.J.^1,2
Department of Physics, Queen's University, Kingston
ON
K7L 3N6, Canada^1
Department of Medical Physics, Cancer Centre of Southeastern Ontario, Kingston General Hospital, Kingston
ON
K7L 5P9, Canada^2
关键词: Attenuation coefficient;    Cone beam;    Converging lens;    Optical attenuation;    Optical CT;    Refractive index mismatch;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/847/1/012007/pdf
DOI  :  10.1088/1742-6596/847/1/012007
来源: IOP
PDF
【 摘 要 】

A series of experiments are conducted to evaluate the effects of refractive index (RI) mismatch in readout with a Vista cone beam optical CT scanner. The changing optical attenuation in cuvettes of different colours and RI are quantified using a spectrophotometer, which shows decreasing attenuation as RI increases. A small jar of a colourless solution inside a large jar of another colourless solution with different combinations of fluids of varying RI is imaged in the Vista scanner. Large RI solutions surrounding small RI solutions create a converging lens, while small RI solutions surrounding large RI solutions create a diverging lens. The same dual-jar system is used with solutions of varying RI and colour (attenuation) to assess the effect of RI on attenuation coefficient determination.

【 预 览 】
附件列表
Files Size Format View
Quantifying refractive index mismatch effects on cone beam optical CT scanner measurements 236KB PDF download
  文献评价指标  
  下载次数:22次 浏览次数:32次