会议论文详细信息
| 14th International Workshop on Slow Positron Beam Techniques & Applications | |
| Evaluation of a positron-beam-pulsing system in KUR reactor-based positron beam facility | |
| Yabuuchi, A.^1 ; Kinomura, A.^1 ; Kuzuya, Y.^1 ; Sato, K.^2 ; Xu, Q.^1 ; Oshima, N.^3 ; O'Rourke, B.E.^3 | |
| Research Reactor Institute, Kyoto University, Kumatori, Osaka | |
| 590-0494, Japan^1 | |
| Department of Mechanical Engineering, Kagoshima University, Kagoshima | |
| 890-0065, Japan^2 | |
| National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology (AIST), Ibaraki, Tsukuba | |
| 305-8568, Japan^3 | |
| 关键词: Beam lines; Electron pulse; Kyoto University Research Reactors; Positron annihilation lifetime measurement; Positron beams; Pulsing system; Slow positron; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/791/1/012013/pdf DOI : 10.1088/1742-6596/791/1/012013 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
The pulsing performance of the positron-beam-pulsing system of the slow positron beamline at Kyoto University research Reactor was tested using electron beams trimmed by 5-mm- or 15-mm-diameter apertures. Electron pulses with full width at half maximum of 143 ps were obtained with a 5-mm-diameter beam which corresponds to the diameter of the brightness-enhanced positron beam. This indicates the pulsing system has a sufficient pulsing performance for positron annihilation lifetime measurements.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Evaluation of a positron-beam-pulsing system in KUR reactor-based positron beam facility | 624KB |
PDF