会议论文详细信息
1st International Symposium 'Physics, Engineering and Technologies for Bio-Medicine' | |
Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine | |
医药卫生;物理学;工业技术 | |
Dubov, L.Yu.^1 ; Belyaev, V.N.^1 ; Berdnikova, A.K.^1 ; Bolozdynia, A.I.^1 ; Akmalova, Yu.A.^1 ; Shtotsky, Yu.V.^1 | |
National Research Nuclear University MEPHI, Moscow, Russia^1 | |
关键词: COMPTON; Compton Camera; Detection efficiency; Field of views; Gamma cameras; Internal diameters; Radionuclide imaging; Spatial resolution; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/784/1/012019/pdf DOI : 10.1088/1742-6596/784/1/012019 |
|
来源: IOP | |
【 摘 要 】
Computer simulations of cylindrical Compton Ar-Xe gamma camera are described in the current report. Detection efficiency of cylindrical Ar-Xe Compton camera with internal diameter of 40 cm is estimated as1-3%that is 10-100 times higher than collimated Anger's camera. It is shown that cylindrical Compton camera can image Tc-99m radiotracer distribution with uniform spatial resolution of 20 mm through the whole field of view.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine | 1090KB | download |