会议论文详细信息
Applications of Novel Scintillators for Research and Industry 2016 | |
A comparison of emerging gamma detector technologies for airborne radiation monitoring | |
物理学;工业技术 | |
Bell, S.J.^1 ; Aitken-Smith, P.^1 ; Beeke, S.^1,2 ; Collins, S.M.^1 ; Regan, P.H.^1,2 ; Shearman, R.^1,2 | |
AIR Division, National Physical Laboratory, Hampton Road, Teddington | |
TW11 0LW, United Kingdom^1 | |
Department of Physics, University of Surrey, Guildford | |
GU2 7XH, United Kingdom^2 | |
关键词: Detection efficiency; Energy resolutions; Gamma detectors; Minimum detectable activities; Monitoring network; Price-performance; Radiation monitoring; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/763/1/012010/pdf DOI : 10.1088/1742-6596/763/1/012010 |
|
学科分类:工业工程学 | |
来源: IOP | |
【 摘 要 】
This paper presents a comparison of new and emerging gamma detector technologies that have the potential to improve in-situ dose and radioactivity-in-air measurements for national monitoring networks. Five detectors were chosen for investigation; LaBr3(Ce), CeBr3, SiPM-CsI(Tl), Cd(Zn)Te and electromechanically-cooled HPGe. These detectors represent the full range of the price-performance matrix. Comparisons have been made of energy resolution, detection efficiency and minimum detectable activity by exposing each detector to a mixed radionuclide source drop-deposited across a filter. Other factors, such as internal radioactivity, linearity, size and cost have also been considered.【 预 览 】
Files | Size | Format | View |
---|---|---|---|
A comparison of emerging gamma detector technologies for airborne radiation monitoring | 1005KB | download |