会议论文详细信息
6th European Conference on Neutron Scattering
A large-area double rotating-crystal monochromator for time-focusing neutron instruments
Zanatta, M.^1 ; Orecchini, A.^1 ; Aisa, S.^2 ; Casinini, F.^2 ; Farnesini, L.^3 ; Deen, P.P.^4 ; Paciaroni, A.^1 ; Petrillo, C.^1 ; Sacchetti, F.^1,2
Dipartimento di Fisica e Geologia, Universitá di Perugia, Perugia, Italy^1
IOM-CNR C/o Dipartimento di Fisica e Geologia, Universitá di Perugia, Perugia, Italy^2
INFN, Sezione di Perugia, Perugia, Italy^3
European Spallation Source, Lund, Sweden^4
关键词: Double rotating;    Key elements;    Long pulse;    Mechanical reliability;    Neutron instruments;    Parallel beams;    Time focusing;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/746/1/012002/pdf
DOI  :  10.1088/1742-6596/746/1/012002
来源: IOP
PDF
【 摘 要 】

We present the principle and the first prototypes of a double rotating-crystal monochromator, based on an assembly of smaller rotating elements. Such a device was developed as the key element to implement a parallel-beam modification of the time-focusing technique for neutron spectrometers. This concept is particularly promising for long-pulse sources and can bring specific advantages on continuous sources as well. Neutron tests performed on the first prototypes validate the mechanical reliability of the proposed design and the feasibility of a large-area double rotating-crystal monochromator based on this technology.

【 预 览 】
附件列表
Files Size Format View
A large-area double rotating-crystal monochromator for time-focusing neutron instruments 2396KB PDF download
  文献评价指标  
  下载次数:8次 浏览次数:13次