Leyton, F.^1,2 ; Ferreira, M.J.^3,4 ; MacEdo, E.M.^3 ; Navarro, V.C.C.^3 ; Garcia, I.F.M.^3,5 ; Pereira, L.L.S.^3 ; Navarro, M.V.T.^3 | |
Faculty of Health and Odontology, Diego Portales University, Santiago, Chile^1 | |
Study Center in Radiological Sciences (CECRAD), Tarapacá University, Arica, Chile^2 | |
Laboratorio de Produtos para A Saúde Do IFBA (LABPROSAUD-IFBA), Bahia, Salvador, Brazil^3 | |
Universidade Estadual da Bahia, Bahia, Salvador, Brazil^4 | |
Bahian School of Medicine and Public Health, Bahia, Salvador, Brazil^5 | |
关键词: Area monitoring; Dose rate; Exposure-time; Radiation calibration; Radiation Exposure; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/733/1/012087/pdf DOI : 10.1088/1742-6596/733/1/012087 |
|
来源: IOP | |
【 摘 要 】
Area monitoring ensures radiation exposure at an acceptable level, which must be lower than the legal limit. Experimental measurements were taken in an ionizing radiation calibration laboratory. The specified reference radiation to radiation protection N80 was used. Five area monitors were used. The ranges of dose rate inaccuracy measured in rate mode for times ≤ 2 and ≥ 3s were from 10 to 48% and from 1 to 15%, respectively. The inaccuracy ranges measured in integrated mode for times ≤ 2 and ≥ 3s were from 4 to 8% and from 0 to 22%, respectively.