| 8th Brazilian Congress on Metrology | |
| Uncertainty evaluation of dead zone of diagnostic ultrasound equipment | |
| Souza, R.M.^1 ; Alvarenga, A.V.^1 ; Braz, D.S.^1 ; Petrella, L.I.^2 ; Costa-Felix, R.P.B.^1 | |
| Laboratory of Ultrasound, Directory of Scientific and Industrial Metrology, National Institute of Metrology, Quality and Technology, Av. Nossa Sra das Graças 50, Duque de Caxias | |
| RJ | |
| 25250-020, Brazil^1 | |
| Institute of Nuclear Sciences Applied to Health, University of Coimbra, Portugal^2 | |
| 关键词: Dead zones; Diagnostic imaging; Diagnostic ultrasound; Guide to the expression of uncertainty in measurements; Measurement uncertainty; Ultrasound images; Uncertainty evaluation; Uncertainty modeling; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/733/1/012043/pdf DOI : 10.1088/1742-6596/733/1/012043 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
This paper presents a model for evaluating measurement uncertainty of a feature used in the assessment of ultrasound images: dead zone. The dead zone was measured by two technicians of the INMETRO's Laboratory of Ultrasound using a phantom and following the standard IEC/TS 61390. The uncertainty model was proposed based on the Guide to the Expression of Uncertainty in Measurement. For the tested equipment, results indicate a dead zone of 1.01 mm, and based on the proposed model, the expanded uncertainty was 0.17 mm. The proposed uncertainty model contributes as a novel way for metrological evaluation of diagnostic imaging by ultrasound.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Uncertainty evaluation of dead zone of diagnostic ultrasound equipment | 1131KB |
PDF