8th Symposium on Frequency Standards and Metrology 2015 | |
Self-generating magnetometer with laser pumping employment in "end resonance" wall coated vapor cell atomic clocks | |
Baranov, A.A.^1 ; Ermak, S.V.^1 ; Smolin, R.V.^1 ; Semenov, V.V.^1 | |
Department of Quantum Electronics, Peter the Great St. Petersburg Polytechnic University, St.-Petersburg, Russia^1 | |
关键词: Analytical studies; Double resonance; Experimental demonstrations; Frequency instabilities; Frequency variation; Microwave frequency discriminators; Microwave transitions; Optically pumped; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/723/1/012016/pdf DOI : 10.1088/1742-6596/723/1/012016 |
|
来源: IOP | |
【 摘 要 】
This paper presents the results of two double resonance signals correlation investigation. These signals were observed synchronously in optically oriented Rb87vapors with laser pumping in a dual scheme: low frequency Mx-magnetometer and microwave frequency discriminator. Analytical studies of the scalar and vector light shift components contribution to the frequency instability of the end resonance microwave transitions are presented. An experimental demonstration of the light shift components mutual compensation in optically pumped Rb87atoms was provided. The results were processed in terms of Allan variance, which demonstrated an effect of decreasing frequency variation at averaging times more than 100 s for a joint scheme of the end resonance microwave transition and selfgenerating (Mx) magnetometer.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Self-generating magnetometer with laser pumping employment in "end resonance" wall coated vapor cell atomic clocks | 1109KB | download |