| 8th Symposium on Frequency Standards and Metrology 2015 | |
| Operation of NIM5 fountain with 1.5x10-15 uncertainty and design of new NIM6 in NIM | |
| Fang, F.^1 ; Liu, N.^1 ; Liu, K.^1 ; Chen, W.^1 ; Suo, R.^1 ; Li, Tianchu^1 | |
| Time and Frequency Division, National Institute of Metrology, Beijing, China^1 | |
| 关键词: Background signals; Cesium fountain primary frequency standard; Cryogenic sapphire oscillators; Density distributions; Microwave synthesizers; Optical molasses; Vacuum system; | |
| Others : https://iopscience.iop.org/article/10.1088/1742-6596/723/1/012009/pdf DOI : 10.1088/1742-6596/723/1/012009 |
|
| 来源: IOP | |
PDF
|
|
【 摘 要 】
The cesium fountain primary frequency standard NIM5 started to operate since 2008 and started to report to BIPM since 2014. The major constrains of NIM5 is a relatively large background signal at the detection and microwave leakages due to the Ramsey cavity. A new fountain clock NIM6 is under construction. Besides some improvements on the vacuum system, a new Ramsey cavity and a microwave synthesizer are made to reduce the Type B uncertainty. Another feature of NIM6 is collecting atoms from a MOT loading optical molasses to get more atoms with a more uniform density distribution. With a new frequency synthesizer based on cryogenic sapphire oscillator (CSO), NIM6 is aiming to reach the quantum projection noise, thus leading to a reduced Type A uncertainty compared with NIM5.
【 预 览 】
| Files | Size | Format | View |
|---|---|---|---|
| Operation of NIM5 fountain with 1.5x10-15 uncertainty and design of new NIM6 in NIM | 1002KB |
PDF