We have developed a multi-channel cross-correlator for high dynamic range (>1010), single-shot temporal contrast measurements. The correlator utilizes a third-order crosscorrelation technique and has a reference channel, to be normalized by the measured peak intensity, and four independent optical delay lines. The measurement results of the shot-to-shot temporal contrast clearly show the intensity fluctuations of short pre-pulses at -4.5 ps and -26 ps before main pulse.