会议论文详细信息
9th International Conference on Inertial Fusion Sciences and Applications
High dynamic range multi-channel cross-correlator for single-shot temporal contrast measurement
Kon, A.^1 ; Nishiuchi, M.^1 ; Kiriyama, H.^1 ; Ogura, K.^1 ; Mori, M.^1 ; Sakaki, H.^1 ; Kando, M.^1 ; Kondo, K.^1
Quantum Beam Science Dirct., Japan Atomic Energy Agency, 8-1 Umemi-dai, Kizugawa, Kyoto
619-0215, Japan^1
关键词: Cross correlation techniques;    Cross correlators;    High dynamic range;    Intensity fluctuations;    Optical delay lines;    Peak intensity;    Reference channels;    Temporal contrast;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/717/1/012103/pdf
DOI  :  10.1088/1742-6596/717/1/012103
来源: IOP
PDF
【 摘 要 】

We have developed a multi-channel cross-correlator for high dynamic range (>1010), single-shot temporal contrast measurements. The correlator utilizes a third-order crosscorrelation technique and has a reference channel, to be normalized by the measured peak intensity, and four independent optical delay lines. The measurement results of the shot-to-shot temporal contrast clearly show the intensity fluctuations of short pre-pulses at -4.5 ps and -26 ps before main pulse.

【 预 览 】
附件列表
Files Size Format View
High dynamic range multi-channel cross-correlator for single-shot temporal contrast measurement 1216KB PDF download
  文献评价指标  
  下载次数:38次 浏览次数:25次