13th South-East Asian Congress of Medical Physics 2015 | |
Model-based assessment of probe placement criteria in cancer therapy using RF ablation | |
物理学;医药卫生 | |
Tanotogono, E.W.^1 ; Akasum, G.F.^1 ; Suprijanto^1 ; Sudirham, J.J.^2 | |
Instrumentation and Control Research Group, Faculty of Industrial Technology, Institute of Technology Bandung, Indonesia^1 | |
Engineering Physics Research Group, Faculty of Industrial Technology, Institute of Technology Bandung, Indonesia^2 | |
关键词: Accumulated damage; Direct observations; Human liver tissues; Minimally invasive; Model-based OPC; Monitoring system; Radio-frequency Ablation; Temperature dependent; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/694/1/012067/pdf DOI : 10.1088/1742-6596/694/1/012067 |
|
学科分类:卫生学 | |
来源: IOP | |
【 摘 要 】
Radiofrequency ablation has been developed as a minimally-invasive method for cancer therapy. Nevertheless, the unfeasibility of direct observation during ablation process sometimes becomes a challenge for practitioners, particularly those constrained by the absence of a proper monitoring system. Thus, aiming to develop a prudent cancer therapy planning, this research develops a 3D model that enable practitioners to predict the tissue damage resulted by a simulated ablation before a real ablation is executed. The model, developed using finite element method, is made to mimic real human liver tissue by simulating its physical properties as temperature-dependent functions. Three probe placement cases, representing three different approaches, are analysed to study the effect of probe placement configuration on tissue damage formed during a time-dependent ablation process. The three placement cases are surface-perpendicular placement, misaligned placement, and relatively accurate placement. It can be concluded that the accuracy of a probe placement configuration can be assessed by quantifying two major parameters: average tissue damage in the target domain and accumulated damage resulted in complementary tissue domain. Optimum ablation duration can also be determined by considering those parameters.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Model-based assessment of probe placement criteria in cancer therapy using RF ablation | 1105KB | download |