会议论文详细信息
International Conference on Particle Physics and Astrophysics
Liquid-xenon detector under the intensive pulse irradiation conditions
Kirsanov, M.A.^1
National Research Nuclear University MEPhI (Moscow Engineering Physics Institute), Kashirskoe highway 31, Moscow
115409, Russia^1
关键词: Absorbed dose;    Liquid xenon;    Pulse irradiation;    Pulsed irradiation;    Stable operation;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/675/4/042014/pdf
DOI  :  10.1088/1742-6596/675/4/042014
来源: IOP
PDF
【 摘 要 】

The effect of intense pulsed irradiation on the operation of the liquid xenon spectrometer was studied. The ionization chamber filled with liquid xenon was irradiated by bremsstrahlung pulses of the microtron. The pulse repetition rate was 400 Hz. The absorbed dose ranged from 10-7to 0.1 Gy per pulse. Stable operation of the liquid xenon spectrometer in the intervals between the pulses of the accelerator was shown for a long time.

【 预 览 】
附件列表
Files Size Format View
Liquid-xenon detector under the intensive pulse irradiation conditions 831KB PDF download
  文献评价指标  
  下载次数:19次 浏览次数:27次