会议论文详细信息
9th International Conference on Modern Techniques of Plasma Diagnostics and their Application
Quadrature laser interferometry in the pulsed plasma diagnostic
Kuznetsov, A.P.^1
National Research Nuclear University MEPhI, Kashirskoe shosse 31, Moscow
115409, Russia^1
关键词: Differential sensitivity;    Linear electron density;    Measurements of;    Pulsed plasma;    Wide dynamic range;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/666/1/012017/pdf
DOI  :  10.1088/1742-6596/666/1/012017
来源: IOP
PDF
【 摘 要 】

The article provides an overview of methods for plasma diagnostics based on the laser interferometry with quadrature recording of informative signals. Quadrature interferometers allow providing a wide dynamic range for measurements of electron densities while maintaining the homogeneous differential sensitivity. The existing range of linear electron density measuring with quadrature interferometers in pulsed plasma installations equals to 1010-1019cm2.

【 预 览 】
附件列表
Files Size Format View
Quadrature laser interferometry in the pulsed plasma diagnostic 1151KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:28次