会议论文详细信息
Electron Microscopy and Analysis Group Conference 2015
Characterisation of Mg biodegradable stents produced by magnetron sputtering
Elmrabet, N.^1 ; Botterill, N.^1 ; Grant, D.M.^1 ; Brown, P.D.^1
Department of Mechanical, Materials and Manufacturing Engineering, Faculty of Engineering, University of Nottingham, University Park, Nottingham
NG7 2RD, United Kingdom^1
关键词: Annealing temperatures;    Biodegradable stents;    Columnar grain structure;    Elevated temperature;    Equi-axed grains;    Hardness and corrosion;    Material densification;    PVD magnetron;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/644/1/012036/pdf
DOI  :  10.1088/1742-6596/644/1/012036
来源: IOP
PDF
【 摘 要 】

Novel Mg-minitubes for biodegradable stent applications have been produced using PVD magnetron sputtering. The minitubes were characterised, as a function of annealing temperature, using a combination of SEM/EDS, XRD and hardness testing. The as-deposited minitubes exhibited columnar grain structures with high levels of porosity. Slight alteration to the crystal structure from columnar to equiaxed grain growth was demonstrated at elevated temperature, along with increased material densification, hardness and corrosion resistance.

【 预 览 】
附件列表
Files Size Format View
Characterisation of Mg biodegradable stents produced by magnetron sputtering 1072KB PDF download
  文献评价指标  
  下载次数:9次 浏览次数:14次