Nounis, C.^1 ; Laoutaris, A.^1 ; Madesis, I.^2,3 ; Dimitriou, A.^2,3 ; Sise, O.^4 ; Benis, E.P.^5 ; Zouros, T.J.M.^2,3
Dept. of Applied Physics, National Technical University of Athens, Athens
GR 15780, Greece^1
Dept. of Physics, University of Crete, Heraklion
GR 71003, Greece^2
Tandem Accelerator Laboratory, INPP, NCSR Demokritos, Ag Paraskevi
GR 15310, Greece^3
Dept. of Science Education, Faculty of Education, Suleyman Demirel Univ., Isparta
32260, Turkey^4
Dept. of Physics, University of Ioannina, Ioannina
GR 45110, Greece^5
关键词: 2D position-sensitive detectors;    Detection plane;    Energy resolutions;    Focal Plane;    Hemispherical deflector analyzer;    Line shape;    Optimum conditions;    Position-Sensitive Detectors;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/635/5/052063/pdf
DOI  :  10.1088/1742-6596/635/5/052063
来源: IOP
PDF
【 摘 要 】

We investigate the energy resolution and line shape of a biased paracentric hemispherical deflector analyzer considering the distance between the exit focal plane and the detection plane. Equipped with a piezoelectrically controlled location of a 2D position sensitive detector we were able to accurately determine the optimum conditions for best energy resolution. The voltages of the injection lens of the spectrometer were also optimized for each location.