Investigation of the dependence of the energy resolution of a hemispherical deflection analyzer on the distance of the position sensitive detector from the focal plane | |
Nounis, C.^1 ; Laoutaris, A.^1 ; Madesis, I.^2,3 ; Dimitriou, A.^2,3 ; Sise, O.^4 ; Benis, E.P.^5 ; Zouros, T.J.M.^2,3 | |
Dept. of Applied Physics, National Technical University of Athens, Athens | |
GR 15780, Greece^1 | |
Dept. of Physics, University of Crete, Heraklion | |
GR 71003, Greece^2 | |
Tandem Accelerator Laboratory, INPP, NCSR Demokritos, Ag Paraskevi | |
GR 15310, Greece^3 | |
Dept. of Science Education, Faculty of Education, Suleyman Demirel Univ., Isparta | |
32260, Turkey^4 | |
Dept. of Physics, University of Ioannina, Ioannina | |
GR 45110, Greece^5 | |
关键词: 2D position-sensitive detectors; Detection plane; Energy resolutions; Focal Plane; Hemispherical deflector analyzer; Line shape; Optimum conditions; Position-Sensitive Detectors; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/635/5/052063/pdf DOI : 10.1088/1742-6596/635/5/052063 |
|
来源: IOP | |
![]() |
【 摘 要 】
We investigate the energy resolution and line shape of a biased paracentric hemispherical deflector analyzer considering the distance between the exit focal plane and the detection plane. Equipped with a piezoelectrically controlled location of a 2D position sensitive detector we were able to accurately determine the optimum conditions for best energy resolution. The voltages of the injection lens of the spectrometer were also optimized for each location.