会议论文详细信息
7th Brazilian Congress on Metrology
Spectral Irradiance Measurements Based on Detector
Lima, M.S.^1 ; Menegotto, T.^1 ; Duarte, I.^1 ; Ferreira Da Silva, T.^1 ; Alves, L.C.^1 ; Alvarenga, A.D.^1 ; Almeida, G.B.^1 ; Teixeira, R.N.^2 ; Couceiro, I.B.^1
Optical Metrology Division, National Institute of Metrology Quality and Technology-INMETRO, Av. N. S. das Graças 50, Duque de Caxias, Xerém RJ
25250-020, Brazil^1
Thermal Metrology Division, National Institute of Metrology Quality and Technology-INMETRO, Av. N. S. das Graças, 50, Duque de Caxias Xerém, RJ
25250020, Brazil^2
关键词: Filter radiometers;    Infrared regions;    Interference filters;    Secondary standard;    Shunt resistances;    Spatial uniformity;    Spectral irradiance;    Spectral irradiance measurements;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/575/1/012023/pdf
DOI  :  10.1088/1742-6596/575/1/012023
来源: IOP
PDF
【 摘 要 】

This paper presents the preliminary results of the realization of absolute spectral irradiance scale at INMETRO in the ultraviolet, visible and infrared regions using filter radiometers as secondary standards. In the construction of these instruments are used, at least, apertures, interference filters and a trap detector. In the assembly of the trap detectors it was necessary to characterize several photocells in spatial uniformity and shunt resistance. All components were calibrated and these results were analyzed to mount the filter radiometer.

【 预 览 】
附件列表
Files Size Format View
Spectral Irradiance Measurements Based on Detector 809KB PDF download
  文献评价指标  