Primary calibration system for vibration transducers from 0.4 Hz to 160 Hz
Ferreira, C.D.^1 ; Ripper, G.P.^1 ; Dias, R.S.^1 ; Teixeira, D.B.^1
National Institute of Metrology Quality and Technology-Inmetro, Division of Acoustics and Vibrations Metrology-Diavi, Vibration Laboratory-Lavib, Av. N. Sra. das Graças 50, Xerém, RJ Duque de Caxias
CEP 25250-020, Brazil^1
关键词: Calibration system;    Data acquisition board;    Frequency ranges;    International standards;    Software for measurements;    Universal counter;    Vibration exciters;    Vibration transducer;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/575/1/012003/pdf
DOI  :  10.1088/1742-6596/575/1/012003
来源: IOP
PDF
【 摘 要 】

This paper presents a system developed at the Vibration Laboratory of Inmetro, which is used for primary calibration of vibration transducers by the fringe counting method. This system includes a vibration exciter, a Michelson interferometer, a data acquisition board, a band-pass filter, a universal counter and a software for measurement automation. It allows the laboratory to perform calibrations in accordance with the international standard ISO 16063-11 in the frequency range from 0.4 Hz to 160 Hz. Some experimental results are presented herein.