会议论文详细信息
8th International Conference on 3D Radiation Dosimetry
Inclusion type radiochromic gel dosimeter for threedimensional dose verification
Usui, Shuji^1 ; Yoshioka, Munenori^1 ; Hayashi, Shin-Ichiro^1 ; Tominaga, Takahiro^1
Department of Clinical Radiology, Hiroshima International University, Higashi-Hiroshima, Hiroshima
739-2695, Japan^1
关键词: Bromoacetic acids;    Dose distributions;    Dose verification;    Highly branched cyclic dextrins;    Inclusion complex;    Leucomalachite green;    Manganese chloride;    Radiation-induced radical;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012035/pdf
DOI  :  10.1088/1742-6596/573/1/012035
来源: IOP
PDF
【 摘 要 】

For the verification of 3D dose distributions in modern radiation therapy, a new inclusion type radiochromic gel detector has been developed. In this gel, a hydrophobic leuco dye (leucomalachite green: LMG) was dissolved in water as an inclusion complex with highly branched cyclic dextrin. The radiation induced radical oxidation property of the LMG gel with various sensitizers was investigated. As a result, the optical dose responses were enhanced by the addition of bromoacetic acid and manganese (II) chloride. Unfavorable auto-oxidation of the gel was reduced when it was stored at 4°C.

【 预 览 】
附件列表
Files Size Format View
Inclusion type radiochromic gel dosimeter for threedimensional dose verification 734KB PDF download
  文献评价指标  
  下载次数:15次 浏览次数:31次