会议论文详细信息
8th International Conference on 3D Radiation Dosimetry
Stray light reduction in optical computed tomography using a convergent cone-beam source
Jordan, Kevin^1,2 ; Dekker, Kurtis^1 ; Battista, Jerry^1,2
London Regional Cancer Program, London Health Sciences Centre, London, United Kingdom^1
Department Medical Biophysics, Western University, London
ON, Canada^2
关键词: Fresnel lens;    Image forming;    Imaging efficiency;    Light level;    Optical computed tomography;    Refractive index matching;    Stray light reduction;    Water solutions;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012033/pdf
DOI  :  10.1088/1742-6596/573/1/012033
来源: IOP
PDF
【 摘 要 】

The planar diffuser light source for a cone-beam optical CT scanner was replaced with a filtered LED and large Fresnel lens. The source was focused on a camera and convergent cone-beam images were acquired. Images are sensitive to mismatches in the refractive index of vessels, samples and matching liquid. For PETE jars and water solutions, a loss of approximately 30% of the projections was demonstrated. Teflon PFA cylinders provided better refractive index matching and more accurate reconstructions. The convergent cone-beam source dramatically increased imaging efficiency by a factor of 1000 and reduced stray light levels by confining illumination to image forming rays.

【 预 览 】
附件列表
Files Size Format View
Stray light reduction in optical computed tomography using a convergent cone-beam source 954KB PDF download
  文献评价指标  
  下载次数:6次 浏览次数:16次