8th International Conference on 3D Radiation Dosimetry | |
Stray light reduction in optical computed tomography using a convergent cone-beam source | |
Jordan, Kevin^1,2 ; Dekker, Kurtis^1 ; Battista, Jerry^1,2 | |
London Regional Cancer Program, London Health Sciences Centre, London, United Kingdom^1 | |
Department Medical Biophysics, Western University, London | |
ON, Canada^2 | |
关键词: Fresnel lens; Image forming; Imaging efficiency; Light level; Optical computed tomography; Refractive index matching; Stray light reduction; Water solutions; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012033/pdf DOI : 10.1088/1742-6596/573/1/012033 |
|
来源: IOP | |
【 摘 要 】
The planar diffuser light source for a cone-beam optical CT scanner was replaced with a filtered LED and large Fresnel lens. The source was focused on a camera and convergent cone-beam images were acquired. Images are sensitive to mismatches in the refractive index of vessels, samples and matching liquid. For PETE jars and water solutions, a loss of approximately 30% of the projections was demonstrated. Teflon PFA cylinders provided better refractive index matching and more accurate reconstructions. The convergent cone-beam source dramatically increased imaging efficiency by a factor of 1000 and reduced stray light levels by confining illumination to image forming rays.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Stray light reduction in optical computed tomography using a convergent cone-beam source | 954KB | download |