会议论文详细信息
8th International Conference on 3D Radiation Dosimetry
Reproducibility assessment of dynamically deforming DEFGEL in a respiratory motion phantom
Franich, R.D.^1 ; Supple, J.R.^1 ; Lindsay, B.^1 ; Yeo, U.J.^1 ; Lonski, P.^1,2 ; Smith, R.L.^1,3 ; Taylor, M.L.^1 ; Dunn, L.^1 ; Kron, T.^1,2
School of Applied Sciences, Health Innovations Research Institute, RMIT University, Melbourne
VIC, Australia^1
Physical Sciences, Peter MacCallum Cancer Centre, Melbourne
VIC, Australia^2
William Buckland Radiotherapy Centre, Alfred Hospital, Melbourne
VIC, Australia^3
关键词: Deformation cycle;    Gel dosimeter;    Reproducibilities;    Respiratory motions;    Setup errors;    Voxel size;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012024/pdf
DOI  :  10.1088/1742-6596/573/1/012024
来源: IOP
PDF
【 摘 要 】

Conducting four-dimensional deforming dosimetry with a deformable gel dosimeter presents several challenges for implementation and interpretation of results. In this work, we describe a modification to a respiratory motion phantom to accommodate the DEFGEL dosimeter, and an evaluation of the reproducibility of the setup and deformation cycle. Setup errors are sub-mm while deformation reproducibility is typically less than half the CT voxel size (1 mm AP and LR; 3 mm SI). 4DCT confirms the absence of hysteresis.

【 预 览 】
附件列表
Files Size Format View
Reproducibility assessment of dynamically deforming DEFGEL in a respiratory motion phantom 851KB PDF download
  文献评价指标  
  下载次数:14次 浏览次数:9次