会议论文详细信息
8th International Conference on 3D Radiation Dosimetry | |
Delivery validation of VMAT stereotactic ablative body radiotherapy at commissioning | |
Olding, T.^1 ; Alexander, K.M.^2 ; Jechel, C.^2 ; Nasr, A.T.^3 ; Joshi, C.^1,2 | |
Cancer Centre of South Eastern Ontario, Kingston General Hospital, 25 King Street West, Kingston | |
ON | |
K7L5P9, Canada^1 | |
Department of Physics, Queen's University, Kingston | |
ON | |
K7L3N6, Canada^2 | |
Department of Chemical Engineering, Queen's University, Kingston | |
ON | |
K7L3N6, Canada^3 | |
关键词: Commissioning process; Crystal violet; Gel measurement; Stereotactic ablative radiotherapies; Volumetric modulated arc therapy; | |
Others : https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012019/pdf DOI : 10.1088/1742-6596/573/1/012019 |
|
来源: IOP | |
【 摘 要 】
Dosimetric validation of two volumetric modulated arc therapy (VMAT) stereotactic ablative radiotherapy (SABR) plans was completed as part of the commissioning process of this technique in our clinic. Static and dynamic ion chamber, EBT3 film and leuco crystal violet (LCV) micelle gel measurements were acquired using a motion phantom with appropriate inserts for each dosimeter. The results show good agreement between measured and calculated plan dose.
【 预 览 】
Files | Size | Format | View |
---|---|---|---|
Delivery validation of VMAT stereotactic ablative body radiotherapy at commissioning | 902KB | download |