【 摘 要 】

Dosimetric validation of two volumetric modulated arc therapy (VMAT) stereotactic ablative radiotherapy (SABR) plans was completed as part of the commissioning process of this technique in our clinic. Static and dynamic ion chamber, EBT3 film and leuco crystal violet (LCV) micelle gel measurements were acquired using a motion phantom with appropriate inserts for each dosimeter. The results show good agreement between measured and calculated plan dose.