8th International Conference on 3D Radiation Dosimetry
Real-time volumetric scintillation dosimetry
Beddar, S.^1
Department of Radiation Physics, University of Texas, MD Anderson Cancer Center, 1515 Holcombe Boulevard, Houston
TX
77030, United States^1
关键词: Detector configuration;    High linear energy transfers;    Image reconstruction methods;    Non-linear response;    Optical corrections;    Plastic scintillator;    Scintillation light;    Scintillator materials;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/573/1/012005/pdf
DOI  :  10.1088/1742-6596/573/1/012005
来源: IOP
PDF
【 摘 要 】

The goal of this brief review is to review the current status of real-time 3D scintillation dosimetry and what has been done so far in this area. The basic concept is to use a large volume of a scintillator material (liquid or solid) to measure or image the dose distributions from external radiation therapy (RT) beams in three dimensions. In this configuration, the scintillator material fulfills the dual role of being the detector and the phantom material in which the measurements are being performed. In this case, dose perturbations caused by the introduction of a detector within a phantom will not be at issue. All the detector configurations that have been conceived to date used a Charge-Coupled Device (CCD) camera to measure the light produced within the scintillator. In order to accurately measure the scintillation light, one must correct for various optical artefacts that arise as the light propagates from the scintillating centers through the optical chain to the CCD chip. Quenching, defined in its simplest form as a nonlinear response to high-linear energy transfer (LET) charged particles, is one of the disadvantages when such systems are used to measure the absorbed dose from high-LET particles such protons. However, correction methods that restore the linear dose response through the whole proton range have been proven to be effective for both liquid and plastic scintillators. Volumetric scintillation dosimetry has the potential to provide fast, high-resolution and accurate 3D imaging of RT dose distributions. Further research is warranted to optimize the necessary image reconstruction methods and optical corrections needed to achieve its full potential.

【 预 览 】
附件列表
Files Size Format View
Real-time volumetric scintillation dosimetry 1806KB PDF download
  文献评价指标  
  下载次数:17次 浏览次数:31次