会议论文详细信息
6th Meeting of the Spanish Neutron Scattering Association
Device and software used to carry out Cyclic Neutron Activation Analysis
Castro-García, M.P.^1 ; Rey-Ronco, M.A.^2 ; Alonso-Sánchez, T.^1
Departamento de Explotación y Prospección de Minas, Universidad de Oviedo, Oviedo
33004, Spain^1
Departamento de Energía, Universidad de Oviedo, Oviedo
33004, Spain^2
关键词: Automatic mechanisms;    Delayed gammas;    Gamma rays detectors;    Multichannel;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/549/1/012006/pdf
DOI  :  10.1088/1742-6596/549/1/012006
来源: IOP
PDF
【 摘 要 】

This paper discusses the device and software used to carry out Cyclic Neutron Activation Analysis (CNAA). The aim of this investigation is defining through this device the fluorite content present on different samples from fluorspar concentration plant through the DGNAA (Delayed Gamma Neutron Activation Analysis) method. This device is made of americium-beryllium neutron source, NaI (2«×2») and BGO (2«×2») gamma rays detectors, multichannel and an automatic mechanism which moves the samples from activation and reading position. This mechanism is controlled by a software which allows moving the samples precisely and in a safe way (∼ms), which it is very useful when the radioactive isotopes have to be detected with a half time less than 8s.

【 预 览 】
附件列表
Files Size Format View
Device and software used to carry out Cyclic Neutron Activation Analysis 742KB PDF download
  文献评价指标  
  下载次数:13次 浏览次数:24次