会议论文详细信息
International Workshop on Neutron Optics and Detectors
Digitized signal acquisition of gas-filled, multiwire, neutron detector with delay-line readout
Orban, J.^1 ; Rosta, L.^1 ; Nagy, A.^2
Wigner Research Centre for Physics, Budapest, Hungary^1
Mirrotron Ltd., Budapest, Hungary^2
关键词: Continuous demand;    Delay line readout;    Detector parameters;    Position resolution;    Position sensitive particle detectors;    Sampling resolution;    Signal acquisitions;    Standard detectors;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/528/1/012044/pdf
DOI  :  10.1088/1742-6596/528/1/012044
来源: IOP
PDF
【 摘 要 】

Multiwire counters are likely to remain a reliable and cost-efficient option for a large class of instruments. Delay-line position encoding is widely used in multiwire position sensitive particle detectors. Improvement of several detector parameters is a continuous demand such as count rate or position resolution. Digitization of the analog detector signals retains all information about the detected events at the highest count rate. Neutronic measurement results on our standard detector with 200mm × 200mm active area are presented with different acquisition algorithms. The analog signals were digitized by a high speed, multichannel Acqiris digitizer with 1 ns sampling resolution.

【 预 览 】
附件列表
Files Size Format View
Digitized signal acquisition of gas-filled, multiwire, neutron detector with delay-line readout 927KB PDF download