会议论文详细信息
17th Pan-American Synchrotron Radiation Instrumentation Conference
Compact pseudo-2D strip detector system for sub-meV IXS
Keister, J.W.^1 ; Siddons, D.P.^1 ; Kuczewski, A.^1 ; Coburn, D.S.^1 ; Beuttenmuller, R.^2 ; Rumaiz, A.^1 ; Cai, Y.Q.^1
Photon Sciences Direc., Brookhaven National Laboratory, Upton, NY 11973-5000, United States^1
Instrumentation Division, Brookhaven National Laboratory, Upton, NY 11973-5000, United States^2
关键词: Detector geometry;    Energy analyzer;    Inelastic X ray scattering;    Sensor geometries;    Silicon devices;    Threshold optimization;    Ultra low noise;    Vacuum enclosure;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/493/1/012015/pdf
DOI  :  10.1088/1742-6596/493/1/012015
来源: IOP
PDF
【 摘 要 】

A dual 32 strip sensor array is realized for use at the IXS (inelastic x-ray scattering) beamline of NSLS-II. By making use of established controls methods and sensor device recipes, our new geometry is realized quickly and at minimal cost. The detector geometry is chosen to match the output of a multi-element high-resolution energy analyzer, while the pulse thresholding is optimized for an ultra-low noise floor at a pass energy of 9.13 keV. Detector subsystems and integration are described, including sensor geometry and silicon device processing, cooling and thermal readback, bias and threshold optimizations, readout ASIC and controls, vacuum enclosure and x-ray window, mounting and positioning, and assembly procedure and testing.

【 预 览 】
附件列表
Files Size Format View
Compact pseudo-2D strip detector system for sub-meV IXS 3529KB PDF download
  文献评价指标  
  下载次数:20次 浏览次数:23次