会议论文详细信息
17th International Conference on the Use of Computers in Radiation Therapy
In vivo and phantom measurements versus Eclipse TPS prediction of near surface dose for SBRT treatments
物理学;计算机科学
Cho, Gwi A.^1,2 ; Ralston, Anna^1 ; Tin, Mo Mo^1 ; Martin, Darren^1 ; Pickard, Sheila^1 ; Kim, J.-H.^2 ; Tse, Regina^1
Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, NSW, Australia^1
Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW, Australia^2
关键词: Film measurement;    Immobilisation;    Monte Carlo calculation;    Phantom measurements;    Stereotactic body radiotherapy;    Treatment planning systems;    Treatment plans;    Virtual water;   
Others  :  https://iopscience.iop.org/article/10.1088/1742-6596/489/1/012008/pdf
DOI  :  10.1088/1742-6596/489/1/012008
学科分类:计算机科学(综合)
来源: IOP
PDF
【 摘 要 】

This study reports on Gafchromic EBT2 film skin dose measurements for lung stereotactic body radiotherapy. These measurements were compared to near surface skin doses predicted by Eclipse treatment planning system (TPS) using the Analytical Anisotropic Algorithm (AAA) for a 6 MV photon beam. The accuracy of the predicted near surface dose for 3×3, 5×5 and 10×10 cm2fields was assessed using an Attix chamber and EBT2 film in a Virtual Water phantom and compared to Monte Carlo calculation. The maximum near surface dose and its location were identified from the patient's treatment plan. For phantom measurements, the TPS dose (nominal 0 mm depth) was higher than the Attix chamber data by up to 24.0 % but in closer agreement with the EBT2 film measurement, which was up to 10.3 % higher than the Attix data. The MC calculated dose values were higher than the Attix data by up to 3.5% for the depths up to 2 mm. The maximum patient skin dose estimated from in vivo EBT2 film measurements was 7.5-19.5 Gy per course and depended on the number of overlapping fields, beam weight and/or contact with immobilisation devices. The TPS predicted dose for patient plans was mostly higher than the in vivo dose, by as much as 69.3%, but in two cases was lower, by as much as -23.1%.

【 预 览 】
附件列表
Files Size Format View
In vivo and phantom measurements versus Eclipse TPS prediction of near surface dose for SBRT treatments 303KB PDF download
  文献评价指标  
  下载次数:25次 浏览次数:41次